User's Manual

(2.7-4.0kPa) lower than the blood pressure measured in the hospital. It is important
to know your normal blood pressure when you are calm at home.
When the cuff is higher than the heart during measurement. If the table or platform
on which the blood pressure monitor is placed too high, causing the cuff to be higher
than the heart, the measured blood pressure value is low.
Have you taken antihypertensive drugs? People who take antihypertensive drugs
will cause high blood pressure when they lose their effectiveness. Please accept
the doctor's guidance.
Is the cuff wrapping position correct? If the cuff position is misplaced, the arterial
signal cannot be captured, and the blood pressure measured is too high. Please
follow the "wind upper arm" page.
Is the cuff wrapping too loose? If the cuff is wrapped too loosely, the compression
force cannot be transmitted to the artery, so the measured blood pressure value is
higher than the actual value. There should be no gap between the arm and the cuff,
please wrap the cuff tightly.
Is the posture correct during measurement? Measuring postures such as bending
over, sitting cross-legged, bending down on a sofa or low table, etc. will cause
abdominal pressure or the position of the arm lower than the heart and raise blood
pressure.
Why is the blood pressure value measured every time different?
Blood pressure fluctuates due to various reasons. Even when measuring at home,
the measurement results will change under the following conditions.
•Within 1 hour after meal, after drinking, drinking coffee, black tea.
•After smoking and bathing.
•After exercise.
•Speaking during the measurement when the mood is irritated due to nervousness
and anxiety.
•When the room temperature changes suddenly when the measurement site or
environment is different from the past.
•After drinking, drinking coffee, black tea.
•After bath.
•After urination.
•When feeling irritable due to nervousness or anxiety.
•When the measurement site or environment is different from the past.
Will it cause blood stasis during continuous measurement? Blood stasis is caused
by pressure on the arm to force the blood to flow through the tip of the finger during
Why is the blood pressure measured at home lower than the blood pressure
measured at the hospital?
We often feel nervous when taking measurements in the hospital. Because we
are in a stable mood when measuring at our home, sometimes it is 20-30mmHg
:NGTQ_U[LUXINUUYOTMU[X9SGXZ(RUUJ6XKYY[XK3UTOZUX
,UXZNKV[XVUYKULVXUVKX[YOTMGTJSGOTZGOTOTMZNK9SGXZ(RUUJ
6XKYY[XK3UTOZUXVRKGYKXKGJZNOYOTYZX[IZOUTIUSVRKZKR_HKLUXKUVKXGZOTM
6RKGYKQKKVZNOYOTYZX[IZOUTSGT[GRYGLKLUXL[Z[XK[YK
/RR[YZXGZOUTYOTZNOYSGT[GRGXKLUXXKLKXKTIKUTR_
 ,GYZKTZNKI[LL]OZNZNKIRUZNH[IQRK
ePA-08B4
User Manual
Smart Blood Pressure Monitor
8.Correct measurement posture
10. Use memory and clear memory function
16. When the screen display error
17. When in doubt
18.Specification
6. Function setting
Entry into static mode
Take measurements with bare
arms or wear light clothing
The center of the cuff and your
heart (chest) should be at the
same height.
9OZ[VXOMNZ
The ideal height difference
between table and chair is
25cm-30cm.
The height difference between
the table and chair in the family
is generally 20-35cm.
measurement. In the event of blood stasis, loosen the cuff, raise your hand over the
top of your head, and repeat the clenching and stretching exercises of about 15
times to clear the stasis.
Throughout the day, even if blood pressure is measured every 10 seconds, there
are changes. In order to manage blood pressure correctly, please measure at the
same time every day.
Why do I feel pain and numbness when the cuff is wrapped around my arm?
This is a temporary phenomenon, so don't worry. When measuring blood
pressure, the cuff needs to be tightened until the blood flow in the arteries is t
emporarily stopped, so you may feel some pain and numbness in the arm. After
removing the cuff, take a short rest to relieve it.
When is the best time to measure blood pressure?
The best time to measure blood pressure is within 1 hour after getting up in the
morning, or before going to bed at night. If you measure in the morning, please take
it within 1 hour after waking up, after urination, and before breakfast (if you are taking
antihypertensive drugs, then before taking the medicine). If the measurement at night
is recommended before going to bed. If it is other time, it is better to measure when
the body and mood are in a stable state. In addition, it is recommended to measure
at the same time every day.
When measurement cannot be performed normally, the following error display will appear.
The posture and cuff winding method shown in the figure below may not be
able to perform correct measurement
The cuff is winding
too loose
ZUURUUYK
9VKGQ
Speak or move during
measurement
Bend down on the
sofa or low table
11.Maintenance and storage
13.Product calibration
Figure1 Figure 2 Figure3
12.Storage method
Degree of protection against harmful ingress of particulate matter
and water
Do not dispose of this product in municipal waste. Collection of
such waste separately for special treatment Is necessary
Serial number
Production Date
The package should be protected from rain
The package contains fragile products, please handle with care
Transport package should be upright
The maximum stacking layer of the same package,
5 means the layer limit
Manufacturer
Refer to instruction manual/booklet
Type BF
The signal is too weak or the
pressure changes suddenly
Too much external
interference information
Please confirm whether the cuff is tied in the
correct way, tightened, and tied in the correct
position
Eliminate the incorrect measurement method
and re-measure with the correct method
When measuring, keep away from equipment
with strong interference sources, such as mobile
phones, motors and other strong interference
sources, and then measure again
Whether to move your arm freely or speak
during the measurement
Display
symbol
Reason/Meaning Correct way
1.Measurement preparation Measurement
1) Wrap the cuff
2) Correct posture
3) Measuring blood pressure
-In order to correct blood pressure management, please measure the blood
pressure at the same time every day.
- Propose to measure your blood pressure an hour after getting up ,or before sleeping
2.Safety information
To assure the correct use of the product, basic safety measures should always
be followed including the note and cautions listed in this instruction manual.
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous.
Follow the instructions of your physician or licensed healthcare provider.
Severe serious blood circulation obstacle, blood disease patients, please
use under the guidance of doctors
This device is intended for use in measuring blood pressure and pulse rate in
the adult population. Do not use this device on infants or persons who cannot
express their intentions.
Please control the pressure within 300mmHg(40kPa).
-Otherwise the arm may be congestion or numbness
This product is only for human body blood pressure measurement.
Please use the special-purpose cuff.
-Otherwise it can’t accurately measure blood pressure
Don’t use the mobile phone near the product .
Don’t dismantle or repair ,renovation the main body of the blood pressure or
cuff.
Use only 1.5V alkaline batteries with this device. Do not use other types of
batteries. This may damage the unit.
If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Contact a physician immediately.
If battery fluid should get in your skin or your clothes, immediately rinse with
plenty of clean water.
Don’t install the wrong battery positive and negative.
When the Low Battery Indicator appears on the display screen turn the
monitor off and remove all the batteries. Replace with four new batteries at the
same time.
Long time no using(more than 3 months) , please take out the batteries.
Dispose of the battery, components and optional accessories according
to applicable local regulations. Unlawful disposal may cause environmental
pollution.
Don’t mix the new and old battery ,or different type of battery.
3.Know your unit
The product are all packed in the gift box, please open the gift box to confirm whether t
he product is complete.
1)Main body
2)Cuff and cuff tube
3)Batteries (4xAA batteries)
4)Instruction Manual
4.Unit Specifications
Main Body
• Before use, please insert four AA batteries into the battery compartment.]]]
4)After the measurement, if the measurement result shows a heartbeat icon to
indicate arrhythmia (such as atrial premature, ventricular premature, and atrial
fibrillation, etc.)
5) The measurement results can be uploaded to the terminal installed with the
corresponding software through wireless transmission.
Software downloading and installing as below:
• For android: Enter into the “Google Play store” to download “Easy BP”
• For IOS: Enter into the “apple store” to download “Easy BP”
• With software installed, click app entering into “login/registration” page. As is
9.Measurement of blood pressure
1)Please keep the correct posture during the measurement, do not move your
body, and keep quiet.
2)In the power-on state, press the [ON/OFF] button, and the blood pressure
monitor starts to inflate automatically. Inflation is stopped, measurement is started
and exhausted. As the cuff vents, the heartbeat icon flashes with each heartbeat.
3)When the measurement is completed, the cuff is completely exhausted. The
blood pressure monitor displays the measurement result, and the measurement
value is automatically stored.
shown in figure 1, if you are a new user, you can click “Join us free” button to
register. As is shown in figure 2, set user information ,then click “done” to finish
register. If you have already registered, please sign in directly.
• The login page is shown as figure 3.
•After finish measurement, there will save the data automatically
Notice: As the software version is updating, different versions may varies in
operations. But the operating methods is generally the same.
6)Remove the cuff and press the [ON/OFF] button to turn off the blood pressure monitor.
•If the user does not understand the measured blood pressure value, please consult
a professional for answers. It is dangerous to make self-judgment and treatment
based on the measurement results. Please be sure to accept the guidance of doctors.
•If you want to terminate the measurement, press and hold the [ON/OFF] button,
the pressurization stops, and the air in the cuff is discharged. The air bag should
not be over-inflated for a long time, otherwise it may cause blood congestion or
danger in the arm.
•Please measure at the same time every day (it is recommended to measure within
1 hour after waking up and before going to bed)
The blood pressure monitor can automatically store measurement results, and can
store 2*100 groups of measurement data. If more than 100 sets of measurement
data have been stored, the oldest set of measurement results will be deleted and
the new measurement results will be recorded.
Do not place the blood pressure monitor in the following places:
•Places exposed to direct sunlight, high temperature, humidity, dust, and
corrosive gases.
•Places that are inclined, vibration, or impact.
•Places where chemicals or corrosive gases are stored.
When not in use for a long time (more than 3 months), please remove the battery
and store it in the box.
In the power-on state, press the [MEM] button to query the stored measurement data.
The default is to query the memory value of "User 1".
If the setting has been changed to "User 2", press the [MEM] button, and the query
is the memory value of "User 2".
In the shutdown state, long press the [MEM] button, when the flashing "M 0", all
memories of the current user will be cleared.
Be sure to observe the items in the manual and other correct usage methods.
If you do not comply, our company will not be responsible for quality.
Maintenance
•Please clean the blood pressure monitor frequently
•If the main body is dirty, please wipe it with a dry soft cloth
•If the main body is extremely dirty, dampen a soft cloth with water or neutral detergent,
wring it out thoroughly, and then wipe the main body.
Maintenance recommendations
•Do not wipe with volatile oil, thinner, gasoline, etc.
•Do not wash or wet the cuff.
•Do not get water to enter the main body.
This product has been calibrated according to the relevant national measurement
requirements when leaving the factory
•It is recommended to check and calibrate the blood pressure monitor once a year
and send it to a metrological verification agency or contact the after-sales service
unit to ensure that the blood pressure monitor functions normally and the
measurement is accurate.
•This product has no user-repairable parts. The company can provide circuit
diagrams, component lists, calibration rules, or information necessary for qualified
technicians to repair.
•If the original parts are replaced with parts not provided by the manufacturer, it
may cause wrong measurement.
14.Product Declaration
•The blood pressure value measured by this product is equivalent to the value
measured by the auscultation method, and its error should meet the requirements
specified in YY0667-2008. For related information, please call the customer
service hotline.
•In order to prevent radio wave interference, this equipment is approved for use in
hospitals, clinical and other environmentally qualified institutions. Using this
equipment in an inappropriate environment may cause electromagnetic interference
to surrounding radio and television equipment. Keep away from strong electric field,
strong magnetic field equipment and high voltage equipment; avoid direct sunlight
to the display screen.
•In order to ensure that the machine can work normally, the user must use it in an
environment with a temperature of +5°C to +40°C and a relative humidity of 15%
to 85%, otherwise the claimed performance specifications may not be met.
•After the product's service life is over, please do not dispose of it as daily garbage.
It should comply with local environmental protection requirements and do proper
disposal. The equipment packaging can be discarded as waste paper for recycling.
15.Blood pressure measurement Q&A
6Set minutes
Press the [SET] button, when the minute value flashes, you can set the minute value.
Each time you press the [MEM] button, it will increase by one. When you keep
pressing the [MEM] button, it will increase quickly.
7)Set unit [default unit is mmHg]
Press the [SET] button to display "0 0" or "0.0 0.0". Long press the [MEM] button to
switch cyclically. The unit of "0 0" is mmHg, and the unit of "0.0 0.0" is KPa.
Note: The switch here is a long press to prevent the unit from being switched by mistake.
Press and hold the [ON/OFF] button to turn on the product, and when it displays
"C-0 0", it will automatically return to zero and enter the static mode.
Note: This operation is recommended not to be used by the user, it is only used for
product verification and maintenance.
7.Wrap the cuff
Air plug
Air plug connector
Please be careful not to press your elbow against the air tube
When winding around the right arm
Make the cuff locate
in the inner arm
•Both left arm and right arm can be measured.
•Take measurements with bare arms or just wearing light clothing.
•Please perform the measurement in a room with a suitable temperature.
•When wearing thick clothes, do not roll up your sleeves for measurement, but take it
off before measuring.
•In order to measure accurately, please
take care to wrap the cuff correctly.
1) Insert the air plug of the cuff into the
air pipe interface of the main body.
2)Thread the right arm into the cuff.The air tube
of the cuff faces the inside of the palm.When
the cuff falls off from the metal sleeve, pass the
most distal end of the cuff through the
D-shapedmetal ring to form a ring. The smooth
cloth surface should beinside the cuff ring.
3)Position the cuff while pulling one end of the
cuff outward.
Cuff
•Applicable arm circumference 22cm36cm
• In order to accurately measure blood pressure, it is recommended to replace it in
time. And your purchased cuff does not include cuff tube, please do not put the
cuff tube away when remove it, and insert it on the new cuff for next use.
• If air leaks due to quality problems, you can call the customer service center to replace it.
• Do not bend the cuff and cuff tube.
•Do not apply pressure before the cuff is wrapped around your arm.
•After turning on the power, you need to set the date and time
•Please set in the order of user, year, month, date, hour, minute, measurement
reminder and unit.
•The blood pressure monitor can automatically store measurement results.
1)Set user [default is user 1]
Press the [SET] button to enter the user selection mode. At this time, the current user
number is displayed, press the [MEM] button to switch users and press the [SET]
button to set users.
2)Set the year value
Display
5.Battery Installation
1)Turn the main body over.
2)Gently press the central serrated part and push in the direction of the arrow to remove
the battery cover.
3) Insert 4 AA batteries as shown in the figure. The positive (+) and negative (-) poles
of the battery should be consistent with the positive and negative poles marked in the
battery compartment.
4) Replace the battery cover.
High pressure
Low pressure
Pulse
Heartbeat/Arrhythmia
User
Warning strip
Number of memory
groups
Time/Date
Low battery symbol
icon "flashing": the battery will run out soon
icon "lights up": the battery is exhausted
5)Set hours
Press the [SET] button, when the minute value flashes, you can
set the minute value. Each time you press the [MEM] button, it
will increase by one. When you keep pressing the [MEM] button,
it will increase quickly.
• Please replace four new batteries at the same time (same type).
•Please turn off the main body power when replacing the battery.
•The stored memory value will not be deleted when the battery is replaced.
•After replacing or removing the battery, the time and date must be reset.
•Please remove the battery if you do not use the product for three months or longer.
•Dispose of the old batteries as per legal regulations.
Long press the [SET] button when the year value flashes, you can
set the year value. Each time you press the [MEM] button,
it will increase by one. When you keep pressing the [MEM] button,
it will increase quickly. Press the [SET] button to set the year value.
The setting range of the year is 2012 to 2099.
3)Set month value
Press the [SET] button, when the month value flashes, you can set
the month value. Each time you press the [MEM] button, it will
increase by one. When you keep pressing the [MEM] button, it will
increase quickly.Press the [SET] button to set the month value.
4)Set the current date
Press the [SET] button, when the date value flashes, you can set
the date value. Each time you press the [MEM] button, it will increase
by one. When you keep pressing the [MEM] button, it will increase
quickly.Press the [SET] button to set the date value.
Inflation error
Abnormal blood pressure
Low battery
Please confirm whether the trachea is folded,
etc., and re-measure after solving
Please confirm whether the tube, tube joints and
the main body are tightly connected, and whether
the tube and cuff are broken.
Then re-measure
Please replace 4 new AA batteries at the
same time
Please rest for 30 minutes before re-measure.
If it is abnormal three times in a row, please go
to the hospital or consult your personal doctor
Reason
Abnormal situation
The measure value
ofblood pressure
abnormal low(high)
No pressure
The cuff exhausts in a
short time
The blood pressure monitor is
working normally, and it is
measured according to the
correct measurement method,
but it is lower (higher) than the
blood pressure measured in
the hospital, and the value
measured each time is different
The cuff is not
wrapped properly
Talk or move arm during
measurement
The rolled up clothes
pressed the arm
The air plug is not correctly
inserted into the air pipe
interface of the main body
Leak in cuff
The cuff is too loose
Taking antihypertensive
drugs.Cuff position
misalignment
Wrap the cuff correctly
Measure quietly.Please refer
to blood pressure
measurement Q&A
Take off the clothes that press
the arm and measure
Connect the air interface and
cuff correctly
Purchase a new cuff.
*The new cuff does not have
an air plug. Please do not throw
away the air plug, install it on a
new cuff and use it
Please wrap your cuff tightly.
*If it is not wound tightly,
excess pressure will be added
to the cuff and its service life
will be shortened.
Please refer to the blood
pressure measurement Q&A.
Handling method
Sudden power failure during
pressurization to reset the
clock
After pressing the button,
nothing is displayed
Other phenomena
If not used for a long time, the
dry battery energy is exhausted
due to temperature changes
Battery energy is completely
consumed
The +-pole of the battery
is reversed
Display screen jitter
Can't measure
Replace with new batteries.
Replace with new batteries.
Insert the battery correctly.
Please press the [ON/OFF]
button and perform the
operation again.
Please try replacing the battery.
Otherwise, it is still unable to
measure, please contact
after-sales service hotline
FCC WARNING: This equipment may generate or use radio frequency energy. Changes or
modifications to this equipment may cause harmful interference unless the modifications are
expressly approved in the instruction manual. The user could lose the authority to operate this
equipment if an unauthorized change or modification is made.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept
any interference received,including interference that may cause undesired operation.
The device has been evaluated to meet general RF exposure requirement.
The device can be used in portable exposure condition without restriction.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules.These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
• Consult the dealer or an experienced radio/TV technician for help.
• The rating information is located at the bottom of the unit.
Product name
Model No.
Measurement methods
Smart Blood Pressure Monitor
ePA-08B4
Oscillometric method
Measuring range
Pressure: 0-280mmHg (0kPa-37.3kPa)
Pulse number: 40 hops/min-199 hops/min
Accuracy
Pressure: within ±3mmHg (±0.4kPa)
Pulse number: accuracy is ±5%
Power supply DC 6V, 4 AA batteries
Operating temperature
+5ȭν+40ȭ15%RHν85%RH,
80Kpa~105Kpa
Storage temperature
-10ȭν+55ȭ10%RHν93%RH,
80Kpa~105Kpa
Dimensions 120 mm (W)*158 mm (L)*54 mm (H)
Accessory
Cuff (applicable arm circumference range
22cm36cm),Instruction manual,Battery
(4*AA batteries), Certificate of Qualification
Shenzhen Belter Health Measurement and Analysis Technology Co., Ltd.
702/704,Block C, Tsinghua Unis Science No.13 Langshan Rd., Hi-Tech Industrial Park(north).
Nanshan District,518057 Shenzhen
PEOPLE’S REPUBLIC OF CHINA
AFTER-SALES SERVICE:
Contact Health Technologies, Inc.
3265 Lawson Blvd.
Oceanside, NY 11572
855-633-5600