Arm Blood Pressure Monitor Instruction Manual Model: AOJ-33B
Table of Contents 1. Unpacking Inspection ...................................... 1 2. Packing List ..................................................... 1 3. Safety Precautions .......................................... 2 4. Product Composition ....................................... 3 5. Intended Use / Instructions for Use ................. 3 6. Contraindication .............................................. 3 7. Product Parts .................................................. 4 8. Preparation ............
Thank you for purchasing the Arm Blood Pressure Monitor. The device uses the oscillometric method of blood pressure measurement. It is intended for professional and domestic use in monitoring diastolic and systolic blood pressure and pulse rate. The device is suitable for use at home, at work, and on vacation. It is also suitable for daily blood pressure checks. 1.
3.
Manufacturer Date of manufacture Serial number LOT Lot number EU authorized representative 4. Product Composition This device is composed of the main body and cuff. 5. Intended Use / Instructions for Use The Arm Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home. 6. Contraindication No known contraindication existing.
7. Product Parts Main Body HD LED Screen Memory Button Start/Stop Button Cuff Type-C Charging Port Speaker Systolic Blood Pressure Icon “Irregular Heartbeat” Indicator Diastolic Blood Pressure Icon Diastolic Blood Pressure Value Systolic Blood Pressure Value 8. Preparation Type-C Charging Please check the device's power before using it. When the battery runs out, please use the manufacturer-provided Type-C Charging Cable to charge the device till the “ flashing.
9. Function Setting Voice On/Off Setting (for voice version only, optional function) In standby mode, press and hold the " and then simultaneously press the " " button, " button for about 3 seconds to enter the voice setting mode. Then press " " button to confirm the selection (as shown in the figure below); " means turn on the sound, " " " means turn off the sound. Voice On Voice Off 10.
--Readings under the conditions listed below may affect results: Within an hour after dinner, after having wine, coffee, tea, sports; talking, being nervous, being in unsteady mood, bending forward, moving, room temperature dramatically changing during measuring; inside a moving vehicle, repeated and continuous measuring. (2) Wearing the Cuff Correctly 1) Unroll cuff. Place your arm inside it. (We strongly recommend to use your left arm.
3) Tighten the cuff around your arm, so it can not move around your arm. Note: Repeated measurement may result in blood congestion in the arm, which will affect the measurement result. To avoid this situation, we advise that you can raise the left hand and hold the fist for several times, or take off the cuff and rest for at least 2-3 minutes before taking the measurement. (3) Measurement Tips • To take a measurement, you need to be relaxed and comfortably seated in a room with a comfortable temperature.
your inner arm at the same level as your heart, with the arm resting comfortably on a table. Note: Please use only the manufacturer-provided cuff, as it has passed all required clinical trials. Any replacement of the original components with materials not specified or provided by the manufacturer may result in measurement errors. (4) Taking a measurement Press the " " button and the monitor will begin to inflate. Please do not move or talk during the measurement.
Note: Consult your doctor if unexpected readings are obtained. (5) Memory function 1) The device can be used for 1 user. This device can store up to 240 sets of measurements. Once the memory log is full, old values are refreshed with new ones. 2) In standby mode, press the " " button once and the device will display the average value of the blood pressure measured for the first three times. Press the " " button again, the first memory will be displayed.
11. Precautions, Warnings and Prompt Instructions • No maintenance or servicing when using. • Do not operate when charging. • Dispose of the monitor when its service life is reached. Follow local regulations regarding the disposal of such products. • Maintenance should be done by the operator as suggested.
user should not self-diagnose or self-medicate per measured results. Kindly adhere to the instructions of your physician or health provider. • DO NOT use this monitor on an injured arm or an arm under medical treatment. • DO NOT use the device while you are on an intravenous drip or blood transfusion. • DO NOT use this monitor in areas containing high frequency (HF) surgical equipment, magnetic resonance imaging (MRI) equipment, computerized tomography (CT) scanners.
• Consult with your physician before using this monitor on an arm where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present because of temporary interference to blood flow which could result in injury. • Consult with your physician before using this monitor if you have had a mastectomy. • Consult with your physician before using this monitor if you have severe blood flow problems or blood disorders as cuff inflation can cause bruising.
necessary, this equipment and the other equipment should be observed to verify that they are operating normally. • Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation. •If continuous measurement, monitor generated heat transits to cuff, it may lead cuff temperature reach about 42℃, lower than 43℃.
home higher than that obtained at the hospital? ● The anti-hypertensive drug the patient might be using has lost its efficacy. Kindly adhere to your doctor's instructions. ● The cuff might not be in the correct position. If the cuff is not placed right, no arterial pressure value will be obtained, and the blood pressure value might be much higher than it is. Therefore, properly position the cuff. ● The cuff is not tight enough.
13. Abnormal Phenomena and Handling If the measurement is abnormal, any of the following symbols may appear. Kindly use the recommended method for measurement. Errors Cause Er U The inflation can not reach 30 mmHg in 12 seconds. Er H The inflation reaches 295mmHg, and it deflates automatically after 20 seconds. Er 1 The pulse rate is not detected correctly. Er 2 Motion or disturbance during measurement. Er 3 Abnormal test result. Er 23 SYS value reads lower than 57mmHg.
Please contact the dealer to replace with Air Leaka- The airbag of the cuff is a new cuff. Don’t ge of Cuff ripped. change the cuff by yourself. If the blood pressure still cannot be measured after trying the above-stated solutions, please contact the dealer. Do NOT attempt to disassemble the device by yourself. 14. Cleaning and Disinfection 14.1 Cleaning The device can be cleaned with a soft, clean cloth dampened with a small amount of neutral detergent or water.
2) The body of the device can also be cleaned with a soft, clean cloth dampened with a small amount of 75% medical-grade alcohol for disinfection. Do not disinfect through methods like hightemperature steam or ultraviolet radiation. These might damage the device and reduce its service life. It is suggested to disinfect the monitor before and after use each time. Each time of disinfection shall be completed within 1min. The number of repeated disinfection each time shall not exceed 2 times. 14.
15. Upkeep and Maintenance water or neutral detergent - Always keep the surface of the device clean and tidy, helpful to prolong its service life. - If the device is dirty, please wipe with a dry soft cloth. If the dirt cannot be eliminated easily, wipe with a soft cloth stained with water or neutral detergent, and then dry with a dry cloth. - No maintenance or maintenance required when operating the device. Do not allow water or other liquids to flow into the device. 16.
Memory 1 user with 240 memory sets Power Source 3.7V Rechargeable Lithium Battery Charging Method Type-C Charging Port; Charging Voltage: DC 5V Charging Time: 2 hours for full charge Low Battery When the power is lower than 3.4V, the device will be turned off. Dimension 125 mm (Length) x 48 mm (Width) x 28 mm (Height) (49.21 inches x 18.9 inches x 11.02 inches) Screen Size 46.5 mm (Length) x 16 mm (Width) (1.8 inches) Cuff Size 22-42 cm (8.66-16.
Transportation and Storage Environment Avoid strong impact, direct impact, exposure or rain during transportation. The device shall be stored indoors at the temperature of -20℃~55℃ and the relative humidity of 10%~93%, atmospheric condition: 70kPa-106kPa without corrosive gas and with good ventilation. 17. Appendix 1 EMC Information Guidance and manufacturer's declaration Electromagnetic emission Arm Blood Pressure Monitor is intended for use in the electromagnetic environment specified below.
Guidance and manufacturer's declaration - Electromagnetic immunity The Arm Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of Arm Blood Pressure Monitor should assure that it is used in such an environment.
Guidance and manufacturer’s declaration - electromagnetic Immunity The Arm Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Arm Blood Pressure Monitor should assure that it is used in such an environment. Radiate d RF IEC6100 0-4-3 (Test specifica tions for ENCLO SURE PORT IMMUNI TY to RF wireless commun ications equipme nt) Test Frequ ency( M Hz) Ba nd (M Hz ) Servic e Modula tion Max .
NOTE: Optional AC adaptor should comply with the requirement of IEC 60601-1 standard. Use only the exclusive AC adapter specified by authorized dealers. Other AC adapter may vary in output voltage and polarities and may represent a risk on your life and damaging the device. When the AC adapter is in use, the device does ot draw power from batteries.
Manufacturing Park, Xiaweiyuan, Gushu Community, Xixiang Street, Bao'an District, 518126 Shenzhen, CHINA Email: info@aojmedical.com Website: https://www.aojmedical.com Tel : 86-755-2778 6026 Made in China Share Info GmbH Heerdter Lohweg 83, 40549 Düsseldorf, Germany Tel:0049 179 5666 508 E-mail: EU-Rep@share-info.com Manual Version: A0 Software version: V1.