Mounting information: The stability of the display must be guaranteed
following mounting of the foot/holder. The immersion depth of the mounting
screws has to be 10 to 12 mm including a 3 mm VESA mounting plate. (See
also table "Mounting screws" on the following). All these requirements are
satisfied when using the original foot. All requirements must be observed
when using customer-specific mounting solutions.
Notice for users: The plug panel closed by the cover (display stand), must not
be opened by users.
Servicing information: If housing components have to be removed for
servicing, this must not be carried out in the presence of patients, the user, or
other persons not involved with servicing.
The following applies to installations in the USA and Canada: Molded power
supply plugs must comply with the requirements for "Hospital Grade
Attachments" UL 498.
Caution
Failure to observe the warnings may result in substantial damage to
property.
Provide sufficient heat dissipation
Slots are provided at the rear of the housing. The display must be placed or
secured on a hard, level surface at least 10 cm from the wall and 15 cm away
from other devices. Several displays can be butt-mounted horizontally and
vertically.
The following must be observed when mounting (VESA connection):
Mounting screws
Number 4
Thread M4
Strength 12
Immersion depth Min. 10 mm; Max. 12 mm
Torque Max. 3 Nm
The permissible ambient temperature range (5 °C ... 40 °C) must not be
violated. Do not subject device to unnecessary shocks. Take care when
transporting! Use the original packaging! The panel in particular should be
protected against shocks.
When touching the panel surface, the mechanical contact or an electrical
discharge may cause a brief disturbance in the picture quality.