5.4 Analog and digital inputs (DVI,VGA,DP, BNC)
DVI socket
With DVI digital signal through DVI cable.
VGA socket
With VGA cable for VGA input.
DP socket
With DP digital signal through DP cable.
BNC connector
Use BNC-VGA cable (optional) for BNC input.
5.5 Power supply connection Note
Device fuses can not be exchanged outside of the repair centers.
The display power supply is connected using an appliance plug. Only use
the power cable supplied in the delivery, or a cable with PE conductor and
appliance socket to DIN 49 547, IEC 320.
Caution
A power cable with PE conductor must be used which corresponds to the
safety requirements of the respective country of use.
5.6 Serial interface
Caution
No other units may be connected to the service socket. Connection or
disconnection of a unit may only be carried out by servicing personnel or
those trained by them. A Serial Spot Meter or Universal Serial Luminance
Meter must not be connected in the presence of patients.
The display has a serial RS 232 6 pins PS2 interface sockets to update
the SW.
5.7 USB interface
The monitor supports touch function ,the touch screen can be used in the
Windows and the Linux..
Figure 4 USB-B connector
Table4
Pin Signal
1 VBUS