4. ENVIRONMENT CONDITONS
4.1 Operation Environment(Philips XN-039059)
Table 4.1
Ambient temperature range
+10 -- +40°C
Ambient humidity 20%-80%
Temperature gradient
<
0.5°C/min
Atmospheric pressure range 70 – 110 kPa
4.2 Transport and Storage (Packed) (Philips XN-039059)
Table 4.2
Ambient temperature range -25 -- +70°C
Ambient humidity 5%-95%
Temperature gradient Max. 10°C/h, no condensation
Atmospheric pressure range 70 – 110 kPa
4.3 Mechanical Load
Operation (Philips XN-039137)
Table 4.3 Vibration
Test direction X-direction
(user position)
Y-direction Z-direction
frequency range
cross-over frequency
10Hz - 150Hz
59.5Hz
10Hz - 150Hz
57.5Hz
10Hz - 150Hz
57.5Hz
acceleration 5g above the cross-over
frequency
2g above the cross-over
frequency
2g above the cross-over
frequency
displacement 0.7mm
peak-peak
up to the
cross-over frequency
0.3mm
peak-peak
up to the
cross-over frequency
0.3mm
peak-peak
up to the
cross-over frequency
sweep rate 1 octave/minute 1 octave/minute 1 octave/minute
initial sweep direction up up up
number of sweeps 10 10 10
Operation (Philips XN-039136)
Table 4.4 Bump
Test direction X-direction
(user position)
Y-direction Z-direction
Acceleration 25g 10g 10g
Pulse width 6 - 10 msec. 6 - 10 msec. 6 - 10 msec.
Number of bumps 4000 1000 1000