User's Manual

Page 23 of 26
Ambient temperature range
+5 -- +35
Temperature gradient
Max. 7/h , no condensation
Relative Humidity 15%-85%
Atmospheric pressure 70 – 106 kPa
Transport and storage (packed)
Ambient temperature range
-20 -- +60
Temperature gradient
Max. 10/h, no condensation
Relative Humidity 10%-90%
Atmospheric pressure 70 – 106 kPa
8.8 Mechanical requirements
Operation
Vibration According to EN 60068-2-6
10 ... 58 Hz with ± 0.075 mm deflection
58 ... 500 Hz at 10 m/s²
Shock According to EN 60068-2-27 (single shock)
150 m/s², 6 ms
No permanent shock allowed in operating conditions
Packed unit
According to 2M2 EN60721-3-2
8.9 Safety specifications
Safety standards
EN60601-1IEC60601-1ANSI/AAMI ES60601-1:2005CSA
C22.2 No.60601-1:2008
Approvals
cTUVusCCCCB (NCB Lab.)
Protection class Protection class I
Degree of protection to
DIN 40050
IP X0
Type B/BF/CF applied
part
No Applied Part
Category AP/APG
equipment
No AP/APG