User's Manual
19" LCD MONITOR
REV A
3
Only use a perfect power supply cable
A damaged power supply cable may result in a fire or electric shock. When
disconnecting the power supply cable, always do so by holding the plug.
Only use the same type of fuse
T2A/250V
Do not insert any objects into the housing
Objects inserted into the housing may result in damage to the unit or personal
injury.
Do not place any objects on top of the units
Penetrating liquids may result in a fire or electric shock.
Connection
No contact to a patient must occur when handling the cables.
Do not hurt yourself, when moving the display
The display can be tilted backwards and forwards. Please, pay attention not to
hurt yourself, when moving the display. Fingers or small objects may get
stuck at the bottom of the display.
When moving the display up and down (height adjustment), make sure you do
not squeeze your hand or any other object. The minimum distance between
the display edge and the bottom is only 100 mm.
Caution
Incorrect installation may result in extensive damage to property.
Installation should be carried out by trained personnel
When installing your medical electrical system with our products in an
environment with patients, please observe the safety requirements of EN
60601-1 (IEC60601-1) for "Specifications for the safety of medical electrical
systems" in order to prevent injury to patients and users of your systems.
Take appropriate measures to particularly ensure that discharge currents
remain below the required limits: Appropriate measures:
- Disconnecting devices for signal input or output unit
- Use of a safety transformer
- Use of additional PE conductor