8.10 Electromagnetic compatibility
EMI voltage/ radiated
interference
EN 60601-1-2
Voltage variations EN 610004-11
Burst on power supply lines
EN 61000-4-4
1 kV
Surge on power supply lines
EN 61000-4-5
1 kV symmetric
2 kV unsymmetric
Static discharge on casing parts
(ESD)
EN 61000-4-2
8 kV air, 4 kV contact
RF irradiation
EN 61000-4-3
80 MHz ... 2,5 GHz
3 V/m 80% AM 1 kHz
Noise immunity
EN 61000-4-6
150 kHz ... 80 MHz
3 V
eff
80% AM 1 kHz
Magnetic constant fields
EN 61000-4-8
Max. 4000 A/m
Magnetic alternating fields
EN 61000-4-8
Min. 10 A/m
Noise emission FCC Part15 B