User's Manual

Last change: 27
th
, Oct. 2005
Copyright @ SSME (SIEMENS
Medical Equipment Ltd.) All
rights reserved
HL1916
Page 22 of 32
5.2 Transport and storage (Packed)
Ambient temperature range
-20 -- +60
Temperature gradient
Max. 5/h, no condensation
Atmospheric pressure 1040 – 674 hPa (0 -- 3048m)
5.3 Mechanical requirements
Operation
Vibration
According to EN60068-2-6
10--58 Hz within ± 0.075 mm deflection
58—500 Hz at 10 m/s2
Shock
According to EN 60068-2-27 (single shock)
150m/s2, 6ms
No permanent shock allowed in operating conditions
Packed unit
Vibration
According to EN60068-2-6
5--9 Hz within ± 3.5 mm deflection
9—500 Hz at 10 m/s2
Shock
According to EN 60068-2-27 (single shock)
250m/s2, 6ms(in storage packing)
According to EN 60068-2-29 (permanent shock)