User's Manual
Table Of Contents
- Customer Specification
- HL1916
- 1. SCOPE
- 2. ELECTRICAL PERFORMANCE
- 2.1 Power Supply
- 2.2 Power Management
- 2.3 Signal Interface
- 2.3.1 Signal Specifications
- 2.3.2 D-SUB Connector and Pin Assignment (Figure1)
- 2.3.3 Digital Visual Interface and Pin Assignment (DVI)
- 2.3.4 Control Interface
- 2.4 Product Features
- 2.5 Screen Performance
- 2.5.1 Standard Testing Conditions
- 2.5.2 Brightness
- 2.5.3 View angle
- 2.5.4 Brightness Uniformity
- 2.5.5 Contrast radio
- 2.5.6 White Color Coordinates
- 2.5.7 Response Time
- 2.5.8 Color Gray
- 2.5.9 Gamma Curve
- 3. OPERATING GUIDE
- 4. MECHANICAL SPECIFICATIONS
- 5. ENVIRONMENT CONDITONS
- 6. DEFECT, SCRATCH and DUST
- 7. NOTICE FOR HANDING
- Appendix 1 Preset Timings
menu using the Set key.
Note: if the OSD menu is quit by changing
the timing or by switching off the monitor,
the modification you have made are saved.
OSD is
locked
If this message is displayed, you do not
have the authority to carry out the changes
in the OSD menu.
Please contact your servicing partner in this
case.
* Service level 2 enter: while the highlight line is on the service level 2 position, press 1 x Up, and 2
x Down keys, it will enter the following submenu.
Service level
2
For LUT select and factory settings recall: Brightness; Contrast; Backlight; Color
3, 6500K; OSD position. (The default state is Gamma CIE)
Gamma CIE Select Gamma CIE LUT
DICOM Select DICOM LUT
Gamma 2.0 Select Gamma 2.0 LUT
Gamma 2.2 Select Gamma 2.2 LUT
Gamma 2.4 Select Gamma 2.4 LUT
Native Select panel native characteristics
4. MECHANICAL SPECIFICATIONS
4.1 Outline dimensions & weight
SIEMENS
Figure 5
Last change: 27
th
, Oct. 2005
Copyright @ SSME (SIEMENS
Medical Equipment Ltd.) All
rights reserved
HL1916
Page 18 of 32