1. SCOPE
This document defines the performance requirements for a 19.0-inch TFT LCD
monochrome monitor for medical use. This product is controlled by model name; any
change will be recorded in the list and confirmed by GE.
This high-resolution monochrome display is specifically designed to meet the
rigorous performance standards needed for diagnostic, interventional radiology, and
other medical applications. To guarantee image integrity, features include accurate
signal conversion and a wide range of interfacing options.
This monitor is factory calibrated to achieve DICOM part 3.14 compliance and
Gamma CIE at the factory set point. The luminance stabilization circuit employs a built
in photo sensor to keep the back-light lamps at a constant luminance for consistent
calibration over the life of the display and can control the back light system
automatically to extend the life of the monitor and achieve very short warming up time.
The surface of the monitor has an anti-glare coating to minimize reflection and a hard
coating to reduce scratch.
With the deliberate designed bracket the monitor can stick on the desktop firmly.
The monitor has a glass on the LCD.(HL1936SMT-L)
The monitor has a touch screen on the LCD. (HL1936SMT-R)