User's Manual

Last change:
2014-5-16
Copyright @ Shenyang
Torch-Bigtide Digital Tech.,Co.,Ltd
HL1706LT
Page 16 of 21
5.3 Mechanical Load
Operation
Vibration
According to EN60068-2-6
10--58 Hz within ± 0.075 mm deflection
58—500 Hz at 10 m/s2
Shock
According to EN 60068-2-27 (single shock)
150m/s2, 6ms
No permanent shock allowed in operating conditions
Packed unit
According to 2M2 EN60721-3-2
6.REGULATION
6.1 Safety Specifications
Safety standards
EN60601-1, EN60601-1-2, ANSI/AAMI ES6060-1: 2005 +
A2(R2012) + A1CAN/CSA-C22.2 NO. 60601-1:14,
FCC PART15B
Approvals
cTUVusCB (NCB Lab.)
Protection class Protection class III
Degree of protection to
DIN 40050
IP 20
Conformity CE
6.2 Electromagnetic Compatibility
IEC60601-1-2 Class B
FCC Part15 class B
6.3 RoHS Compliance
Comply with RoHS Directive 2011/65/EU
7. MTBF & Warranty
The monitor life time: 50,000 hours (Brightness reduction to 50%)
Warranty items refer to related service contract.
8. DEFECT, SCRATCH and DUST
8.1 Inspection instruments
8.1.1 Pattern generator: LD-2000 or equivalent model.
8.1.2 Video board: AU video board or equivalent. The output of the signal should comply with the
specification provided by AU.