User Manual

Shenchen Pump AMC Series
- 6 -
5. Function Instruction
Adjustment function of pressure tubing gap
Adjust the knob to adjust the pressure tubing gap, in this way you can change the
pump head and tubing outlet pressure and flow rate.
The pressure tubing gap is stepless adjustment, any position within the adjustment
range can be used normally.
Pump head pressure standard:
1) If the pump head will be used for 2*1 silicone tubing, the pressure is 0.2±
0.03Mpa.
2) If the pump head will be used for other size tubing, adjust the pressure
according with the specific tubing size.
Turn the knob clockwise, the position of the knob becomes lower, the pressure
tubing gap is reduced accordingly. And if the pump head tubing outlet pressure
increases, the flow rate becomes lower.
Turn the knob counterclockwise, the position of the knob becomes higher, the
pressure tubing gap is increased accordingly. And id the pump head tubing outlet