User Manual
Table Of Contents
- Please read the manual carefully before operating the product.
- Tubing may have cracks due to wear, which results in overflow of fluid from the tubing. In that time human body and instruments may be harmed. So, users must check frequently and change the tubing in time.
- Please stop pump head rotating when replacing the tube or changing the tube position, otherwise it may cause fingers or clothes got caught.
- There are moving parts of the pump head inner, before opening the upper block, it must be carried out according to the following requirements:
- Ensure the pump is isolated from the main power supply.
- Ensure there is no pressure in the tubing.
- If the tubing has defects, ensure the liquid in the tubing is discharged to other containers or drain tubing.
- If transferring dangerous liquid, the user must wear protective clothing and eye protection.
- 1. Product Introduction
- 2. Pump Head Instruction
- 3. Dimension Drawing
- 4. Pump Head Technical Specification
- 5. Function Instruction
- 6. Troubleshooting
- 7. Maintenance
- 8. Warranty and After Service
Shenchen Pump AMC Series
- 6 -
5. Function Instruction
Adjustment function of pressure tubing gap
Adjust the knob to adjust the pressure tubing gap, in this way you can change the
pump head and tubing outlet pressure and flow rate.
The pressure tubing gap is stepless adjustment, any position within the adjustment
range can be used normally.
Pump head pressure standard:
1) If the pump head will be used for 2*1 silicone tubing, the pressure is 0.2±
0.03Mpa.
2) If the pump head will be used for other size tubing, adjust the pressure
according with the specific tubing size.
Turn the knob clockwise, the position of the knob becomes lower, the pressure
tubing gap is reduced accordingly. And if the pump head tubing outlet pressure
increases, the flow rate becomes lower.
Turn the knob counterclockwise, the position of the knob becomes higher, the
pressure tubing gap is increased accordingly. And id the pump head tubing outlet