- Tubing may have cracks due to wear, which results in overflow of fluid from the tubing. In that time human body and instruments may be harmed. So, users must check frequently and change the tubing in time.
- Please stop pump head rotating when replacing the tube or changing the tube position, otherwise it may cause fingers or clothes got caught.
- There are moving parts of the pump head inner, before opening the upper block, it must be carried out according to the following requirements:
- Ensure the pump is isolated from the main power supply.
- Ensure there is no pressure in the tubing.
- If the tubing has defects, ensure the liquid in the tubing is discharged to other containers or drain tubing.
- If transferring dangerous liquid, the user must wear protective clothing and eye protection.
- 1. Product Introduction
- 2. Pump Head Instruction
- 3. Dimension Drawing
- 4. Pump Head Technical Specification
- 5. Function Instruction
- 6. Troubleshooting
- 7. Maintenance
- 8. Warranty and After Service
Shenchen Pump AMC Series
- 7 -
pressure is reduced, the flow rate become higher.
Note:
The lower the knob, the greater the outlet pressure of the pump head tubing, the
faster the tubing wears, and the shorter the tubing life. Therefore, it is not
recommended to adjust the knob too low.
Note:
1. The knob is lower, the outlet pressure on pump head tubing is greater. The faster
the tube will wear, and the shorter the tube life will be. So it is not recommended to
set the knob too low.
2. The higher the knob, the outlet pressure of the pump head tubing is lower. And the
lighter the tubing is compressed, and the easier it is to cause the problems with no
liquid or dripping from at the outlet side.
6.
Troubleshooting
If the pump head is running but the flow is small or no flow, please check the
following items:
Check whether the pump is supplied with transmission liquid.
Detect whether the suction side pipeline of the pump is blocked by suction
wall.
Check if the tube is cracked.
Check for blockages or kinks in the tube.
Check whether the wall thickness of the tube used is correct.
Check whether all valves in the tube are open.
Check whether the rotation direction of the pump is correct.
7. Maintenance
When pump does not work for a long time, please loosen the cartridges from
pressing on the tubing to avoid changing the shape of tubing because of
longtime extrusion.
Keep the rollers of pump head clean and dry, otherwise it can quicken the