SSI5R-HS SSI5R-HS-2 Orbital Shaking Incubator Operational Manual 4861640 Revised 12/2013 Formerly designated as SI6R-HS SI6R-HS-2
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Table of Contents Interior Accessory Outlet ........................................................................................... 12 Maintaining the Unit 13 Disinfecting Your Unit .............................................................................................. 13 A Typical Decontamination Procedure ...................................................................... 13 Control Maintenance ........................
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Using the Unit Safely Introduction Thank you for choosing a SHEL LAB shaking orbital incubator. SHEL LAB sets the standard for quality and reliability. Your unit is backed by over 30 years of design and manufacturing excellence in the scientific, research, and medical equipment industries.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Meanings of Symbols About this Manual In this manual and on labels attached to the product, graphic symbols have the following meanings: Throughout this manual, the words WARNING and CAUTION have the following meanings: Symbol Indcation You should consult this manual for a description or discussion of a control or item Temperature WARNING A potentially hazardous situation that, if not avoided, could result in serious injury or deat
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Features of Your Unit Product Description Key Features Your shaking orbital incubator provides: Controlled environment For continuous growth of biological organisms. Vibration-free operation A unique adjustable counterbalance system provides vibration-free operation regardless of load. Large chamber A large six cubic foot chamber facilitates throughput.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Receiving Your Unit Unpacking Your Unit and Inspecting Before leaving our factory, all units are packaged in high quality shipping materials designed to provide protection from transportation related damage. Once a unit leaves our factory, however, safe delivery becomes the responsibility of the carrier who is liable for loss or damage to your unit. Damage sustained during transit is not covered under your unit warranty.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Installing the Unit that the door can swing opens at least 90 degrees. Do not block access to the Installation Overview power cord, circuit breaker or fuses. To install your unit, you need to: 1. Select a suitable operating location for the unit. 2. Level the unit. 3. Sterilize the unit. 4. Install the sample tray. 5. Plug the unit into a power source.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual sterilizing your unit, see “Disinfecting Your Unit“ on page 14. Installing Sample Tray Your unit comes with a sample tray as standard equipment. To install the sample tray 1. Enclose all corners of the shaking mechanism within the lips of the sample tray. This can be done easily by positioning the front two corners and then setting the rest of the tray down. 2. Shake the tray by its handles to confirm that it is firmly in place.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Operating the Unit Getting the Unit Ready for Use Control Panel Overview Before turning the incubator on for the first time, take a moment to familiarize yourself with its controls and features. Following is an overview of the control panel. WARNING This equipment is NOT intended for the processing of Flammable materials. 1. Main temperature control 2. 3.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual the samples. Be sure the thermometer is not touching any shelving. Turning the Unit On The unit is equipped with an On/OFF switch that controls power to the entire unit. The switch is lit by a green light when the power is on. To turn the unit on 1. Be sure the unit is plugged in. 2. Push the Power switch to the On (I) position. 3.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual To set the safety temperature thermostat Using the Timer 1. Using the incubator’s timer, you can run the shaker platform at a preset RPM for a preset time. The timer can be set at intervals of one (1) minute up to a maximum of a 999 (16 hours 39 minutes). For best results, calibrate your unit before you set the safety thermostat. See “Calibrating the Main Temperature Control” on page 11.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Interior Accessory Outlet This unit features an accessory outlet to provide power for equipment such as magnetic stirrers, rockers, etc. The weight of this equipment should not exceed 25 pounds (55.125 kg) per shelf. This equipment may provide additional heat that could affect the temperature range of this incubator. We recommend testing the incubator and any accessory equipment to ensure that the desired operating conditions can be met.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Maintaining the Unit The only regular maintenance required for your unit is to keep it clean and free from contamination. Use the guidelines and instructions in this section to maximize the life of your incubator and help prevent contamination of your samples. Disinfect the parts with a 70% alcohol solution. Rinse with distilled water and wipe dry with a soft cloth.
SHEL SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Troubleshooting Solving Problems WARNING Should the proper function of your unit come into question, use this section to help you determine what the problem is and how to fix it. Replacing fuses with wrong type and value can result in serious damage to the equipment and property. ONLY Replace fuses with the same type and amperage indicated on the replacement fuse labels. Check if your question is similar to those listed below.
SHEL SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual What is the problem? Possible Causes To solve the problem... problem... Call customer service. The Main Temperature Control displays “LO”. The unit will not heat up to set point. Call customer service. Probe has shorted out. The amperage and voltage of the unit’s power source do not match the unit’s requirements. 1. - Or - The unit will not heat at all. Over Temperature Protection (OTP) is set too low.
SHEL SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual What is the problem? Ice built up in the chamber. Possible Causes The door gasket leaks. The door is opened too often. There’s an open container letting moisture collect inside the chamber. To solve the problem... 1. 2. 3. Check door seal. Try to limit door opening/closing. Seal the container. Power The unit will not turn on. Power cord not firmly plugged into the outlet.
SHEL SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual What is the problem? Possible Causes To solve the problem... tightened enough. If this does not solve the problem... - Or - The hinges are not adjusted properly. The door has been twisted. - Or - The unit has been damaged and the body is not square. 2. - Or - The shaker motor squeals continuously with a constant pitch. Changes in intensity only when rpm varies. Stops when the oscillate switch is turned off.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Getting Your Unit Serviced Getting Assistance Returning Your Unit Your incubator will provide years of trouble-free operation. Should you require assistance, however, SHELLAB customer service and support is available to assist you. If you need to return your unit for any reason, first contact your customer representative for return authorization number (RA#). Be sure to print the RA# clearly on the package in which you ship your unit.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Replacement Parts and Accessories Replacement Parts Part 115V 220V Adjustable feet 2700500 2700500 Motor, Circulation Alarm Display Vertical 1750681 1750681 Outlet, Interior Counterweight, Single 5071329 5071329 Platform (Sample Tray) 9750758 9750758 Door Gasket 3450562 3450562 Power Cord 1800516 1800537 Element Coil 9570703 9570738 Refrigeration Unit 7010521 7010541 Flask Clamps, 1 Liter 9530532 9530531 S
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual Specifications All Models Temperature Unit Range 10C (5°C min.) above Ambient to 60C Uniformity +0.8 C at 37 C Accuracy +0.1 C Alarms Visual Safety Lamps Capacity Volume 156 m3 (5.5 cu. ft.) Shelves Supplied 2 stainless steel Shelf Dimensions 47 × 47 cm (18.5 × 18.5 in.) Total Shelf Capacity 10 kg (22 lbs) Dimensions Interior (Width × Depth × Height) 48.3 × 48.9 × 59.7 cm (19 × 19.25 × 23.5 in.
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual WIRE DIAGRAM SSI5R-HS (SI6R-HS) 110V (9851461) 2800502 EMI FILTER BLUE INLET 4200505 BROWN BROWN BLUE POWER SWITCH 7850533 BROWN BLACK/VIOLET BLACK/WHITE BLACK BLACK/YELLOW 120V / 240V TO 90V STEP DOWN TRANFORMER 8350509 TIMER SWITCH 7850553 YELLOW YELLOW J11 MOTOR RED BROWN ORANGE MOTOR N ORANGE TEMP DISPLAY 1750791 J12 MAIN CONTROL BOARD 1750787 RPM DISPLAY 1750790 DOOR SWITCH 7850578 LOAD HOT VALVE HEATER J10
SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual SSI5R-HS-2 (SI6R-HS-2) 220V (9851462) 2800502 EMI FILTER BLUE BLUE INLET 4200505 BROWN BROWN BLUE BLACK/WHITE BLACK 120V / 240V TO 90V STEP DOWN TRANFORMER 8350509 BLACK/VIOLET BROWN POWER SWITCH 7850533 BLACK/YELLOW TIMER SWITCH 7850553 YELLOW YELLOW J11 MOTOR RED BROWN ORANGE MOTOR N ORANGE TEMP DISPLAY 1750791 J12 MAIN CONTROL BOARD 1750787 RPM DISPLAY 1750790 DOOR SWITCH 7850578 LOAD HOT VALVE HEATER J10 MO