DROSOPHILA SPECIFIC INCUBATORS SRI21F SRI21F-2 SRI21FV SRI21FV-2 Previously designated LIFLY LIFLY-2 LIFLY-VIEW LIFLY-2VIEW Installation and operations manual Revised 5/2014 4861614 Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113 EMAIL: tech@Shellab.com INTERNET: http://www.Shellab.
TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 INSTALLATION SECTION 3.0 GRAPHIC SYMBOLS SECTION 4.0 CONTROLS OVERVIEW SECTION 5.0 OPERATION SECTION 6.0 MAINTENANCE SECTION 7.0 TROUBLESHOOTING SECTION 8.
1 Section RECEIVING AND INSPECTION Your satisfaction and safety require a complete understanding of this unit. Read the instructions thoroughly and be sure all operators are given adequate training before attempting to put the unit in service. NOTE: This equipment must be used only for its intended application; any alterations or modifications will void your warranty. 1.1 Inspection: The carrier, when accepting shipment, also accepts responsibility for safe delivery and is liable for loss or damage.
2 Section INSTALLATION This unit should remain upright for 24 hours prior to operating to allow the oil in the refrigeration compressor to settle. Local city, county or other ordinances may govern the use of this equipment. If you have any questions about local requirements please contact the appropriate local agency. Installation may be performed by the end user.
3 Section GRAPHIC SYMBOLS Your incubator is provided with a display of graphic symbols on the control panel which are designed to help identify the use and function of the adjustable components. 1. Indicates that you should consult your manual for further description and discussion of a control or user item. 2. Indicates “Temperature” 3. Indicates “Overtemperature” 4. C Indicates “Degrees Centigrade” 5. Indicates “AC Power” 6. Indicates “Manual Adjustment” 7.
4 Section CONTROLS OVERVIEW CONTROLLER KEYS AND DISPLAYS INFINITE KEY ALLOWS TO BACK UP ONE LEVEL OR RETURN TO HOME PAGE EASY ZONE KEY 1 STARTS AND STOPS PROGRAM 20.0 ACTUAL PROCESS TEMPERATURE DISPLAY C EZ1 20.
5 Section OPERATION The refrigeration system, heater, and air circulating fan are used in conjunction with the temperature control circuit to achieve sensitive temperature control. The thermostat sensor located in the air stream senses any temperature deviation from the control point, and heat is provided to maintain desired temperature. The circulating fan provides even air distribution throughout the chamber and assures temperature uniformity.
PRESS AND HOLD BOTH UP AND DOWN USE UP OR DOWN ARROWS 20.0 20.0 A LooP OPer OPer PRESS INFINTE KEY REPEATEDLY -5.0 UNTIL NEXT SCREEN DISPLAY APPEARS TO SELECT NEXT SCREEN DISPLAY ARROWS FOR 3 SECONDS OR UNTIL NEXT SCREEN PUSH ADVANCE KEY REPEATEDLY USE UP OR DOWN ARROWS 0.2 db TO CHANGE PARAMETER O.2 TO -5.0 20.0 20.0 db TO RETURN TO HOME PAGE Changing the parameter from 0.2 to -5.
5.11 Exterior Heat: Under normal operating conditions the unit will generate enough heat to be felt by the hand when touching the sides of the unit. This is normal and does not indicate improper performance. 5.12 Programming Controller for Light Operation: PROGRAMING THE DAY AND NIGHT CYCLE PROGRAMING THE DAY AND NIGHT CYCLES ARE DONE IN JUST 8 EASY STEPS. THE FIRST TWO STEPS ARE ENTERING THE REAL TIME IN HOURS AND MINUTES.
5.13 Operating Lights Manually Without Running Program: To turn lights on without running a program ENT1 (event output 1) must be turned on. Follow instruction below on how to turn the output on. PRESS AND HOLD BOTH UP AND DOWN USE UP OR DOWN ARROWS A 20.0 20.0 P.
F.L 0.5 t.tun no AC.LF 60 lEr OFF A.tsp 90 Rtyp t1 dEL 0.0 t.agr Crit P.tyt StPt i.ca 0.0 P.dL 5.0 gSE OFF A.in ambient U.FA USEr SiA 5 i.Er none FAIL USEr Sib 6 L.dE No Pot1 9999 rP OFF C.Led both L.SP 0.0 ZonE on h.SP 45 ChAn on C.SP 20.0 d.Prs 1 rtC Hour Min dow Current time hours Current time minutes Day of week l.dS AMPIENT d.t1 0 SP.lo 100 USr.S none 1 SP.hi 100 USr.r none map 1 0.SP 0.00 NUS YES C.
6 Section MAINTENANCE The design of the chamber is such that periodic maintenance is kept to a minimum. NO lubrication or adjustments of components is needed. If the incubator is used frequently at temperatures below ambient room temperature or in any manner that increases moisture build-up within the chamber, a frequent defrosting schedule is recommended. 6.1 Defrosting: Frost can appear inside the unit due to moisture accumulating and condensing on the coldest surface.
7 Section TROUBLESHOOTING and SERVICE When troubleshooting, always make a visual inspection of the incubator and its control console to find loose or disconnected wires which may be source of the trouble. In the event the incubator does not operate properly, check the following before calling for service. TEMPERATURE Temperature too high 1/ controller set too high-see section 5.
Unit won't cool 1/ be sure that fan is circulating air in the chamber and over the compressor. 2/ confirm proper sensor location and operation. 3/ look for leaks in the chamber or around the door gasket. 4/ assure ample room around the unit as described in Installation section 2.2. 5/ compare ambient specifications to Unit specifications in section 8.0. Ice build up in chamber 1/ Search for leak in door gasket. 2/ door being opened too often. 3/ open container inside the chamber.
8 Section PARTS LIST Description Blower Motor Circuit Breaker Convenience Outlet Defrost Switch High Limit Control I/O (on/off) Switch Low Limit Control Temperature Controller Watlow PM Overtemperature Thermostat Pilot Light, Green Pilot Light, Red Relay Mechanical Relay Solid State RTD Temperature Probe Fan Guard Power Cord 100-120V 220-240V Units Units 4880564 1100500 6100525 7850579 1750538 7850553 1750538 9660515 1750862 4650554 4650553 7030536 7030533 6600520 2600518 1800529 4880564 1100500 6100531
UNIT SPECIFICATIONS Shipping Weight SRI21F (LIFLY) & SRI21FV (LIFLYVIEW ) 340 lbs. SRI21F-2 (LIFLY-2) & SRI21FV-2 (LIFLY2VIEW) 340 lbs. Net Weight 246 lbs. 246 lbs. Exterior WxDxH (in.) Dimensions 32 x 32 x 77 32 x 32 x 77 Interior WxDxH (in.) Dimensions 27 x 23.5 x 57 27 x 23.5 x 57 Capacity 20.3 cubic feet 20.3 cubic feet Capacity 305 bottles 305 bottles Temperature Range 0° TO 45°C 0° TO 45°C Electrical 100-120V 9.5 220-240V 4.0 The equipment is rated for ingress protection IPX0.
WIRING DIAGRAM SRI21F (LIFLY) SRI21FV (LIFLY-VIEW) 100V-120V 9851583 17
WIRING DIAGRAM SRI21F-2 (LIFLY-2) SRI21FV-2 (LIFLY-2VIEW) 220-240V 9851584 18