ECONOMY INCUBATOR SMI1EM SMI1EM-2 Previously designated as: EI1M EI1M-2 21/2013 4861654 INSTALLATION AND OPERATIONAL MANUAL Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113 EMAIL: tech@Shellab.com INTERNET: http://www.Shellab.
TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 GRAPHIC SYMBOLS SECTION 3.0 INSTALLATION SECTION 4.0 CONTROL PANEL OVERVIEW SECTION 5.0 OPERATION SECTION 6.0 MAINTENANCE SECTION 7.0 TROUBLESHOOTING SECTION 8.
1 Section RECEIVING AND INSPECTION Your satisfaction and safety require a complete understanding of this unit. Read the instructions thoroughly and be sure all operators are given adequate training before attempting to put the unit in service. NOTE: This equipment must be used only for its intended application; any alterations or modifications will void your warranty. 1.1 Inspection: The carrier, when accepting shipment, also accepts responsibility for safe delivery and is liable for loss or damage.
2 Section GRAPHIC SYMBOLS Your incubator is provided with a display of graphic symbols to help in identifying the use and function of the available adjustable components. 2.1 This symbol, when shown, indicates that you should consult your manual for further description or discussion of a control or item. 2.2 Indicates “AC Power” 2.3 Indicates “Heating” 2.4 Indicates ”Adjustable Temperature” 2.5 Indicates “Manual Control” 2.6 Indicates “Earth Ground Symbol” 2.
3 Section INSTALLATION Local city, county or other ordinances may govern the use of this equipment. If you have any questions about local requirements, please contact the appropriate local agency. Installation may be performed by the end user. Under normal circumstances this unit is intended for use indoors, at room temperatures between 5 above ambient to 70C, at no greater than 80% Relative Humidity (at 25C) and with a supply voltage that does not vary by more than 10%.
3.6 Shelf Placement: Place the two shelves in the desired position. See Figure 1.
4 Section CONTROL PANEL OVERVIEW 4.1 Power Switch: The main power I/O (on/off) switch controls all power to the unit and must be in the I or ON position before any systems are operational. 4.2 Temperature Controller: This control is marked SET TEMPERATURE and consists of the digital display and Up/Down arrow pads for inputting set point temperature and calibration. 4.
5 Section OPERATION 5.1 Check power supply against unit serial plate; they must match. Plug service cord into the grounded electrical outlet and turn the power switch ON. 5.2 Thermometer Placement: Place the thermometer and clip (provided in the accessory kit) through the hole at the top of incubator. The thermometer is to verify operating temperature. See Figure 3. 5.3 Set Temperature Controller: To set the main temperature controller, perform the following steps: 1. 2. 3. 4. 5.
Figure 3
6 Section MAINTENANCE NOTE: Prior to any maintenance or service on this unit, disconnect service cord from the power supply. 6.1 Cleaning: Clean interior of the incubator on a regular basis. Remove shelves and shelf clips and sterilize the incubator with a disinfectant that is appropriate for your application. The shelves and clips are autoclavable, or can be cleaned with the same solution as the incubator.
7 Section TROUBLESHOOTING FOR PERSONAL SAFETY, ALWAYS DISCONNECT THE POWER BEFORE SERVICING. Always make a visual inspection of the incubator and control panel when troubleshooting. Look for loose or disconnected wires that may be the source of the trouble. TEMPERATURE Temperature too high 1/ controller set too high. 2/ controller failed on – call Customer Service Temperature too low 1/ controller set too low. 2/ unit not recovered from door opening – wait for heating indicator to turn off.
8 Section PARTS LIST Description 115V Circuit Breaker Heating Element I/O Switch Power Cord, USA Temperature Controller 1100505 2350562 7850570 1800529 1750549 UNIT SPECIFICATIONS Unit Dimensions WxDxH Exterior Interior SMI1EM & SMI1EM-2 16.75x17.75x22.25 (EI1M & EI1M-2) 12x12x14 Weight Net Shipping 53 lbs. 44 lbs. Capacity Cubic Ft Temperature C 1.0 Amb.
WIRE DIAGRAM SMI1EM (EI1M) (9851440) POWER CORD 1800529 CURCUIT BREAKER 1100505 GREEN LIGHTED SWITCH 7850570 WHITE HT SENSOR HOT NEUT RA L GROUND TEMPERATURE CONTROL 1750986 LOA D 2350562 120V 250W
WIRE DIAGRAM SMI1EM-2 (EI1M-2) (9851441) POWER CORD 1800537 GREEN LIGHTED SWITCH 7850570 WHITE HT SENSOR HOT NEUTRAL GROUND TEMPERATURE CONTROL 1750987 LOAD 2350500 120V 125W 2350500 120V 125W