Owner's manual

13
1/ ±0.1 may be normal.
2/ For G suffix models: may vary ±2.0 degrees.
3/ For F suffix models: is fan working? verify movement of air in
chamber.
4/ Is ambient room temperature radically changing either door
opening or room airflow from heaters or air conditioning ? stabilize
ambient conditions.
5/ This may happen if exhaust stack is 100% open or if power exhaust
is cycling adjust stack to at least ¼ closed.
6/ Sensor miss-located, damaged or wires may be damaged - check
mounts for control and OTP sensors, then trace wires or tubing
between sensors and controls.
7/ Calibration sensitivity call Customer Service.
8/ High limit set too low be sure that Safety Thermostat is set more
than 5 degrees over Main Controller set point; check if OTP pilot is on
continuously; turn controller knob completely clockwise to see if
problem solved then follow instructions in section 6.5 for correct setting.
9/ Electrical noise remove nearby sources of RFI including motors,
arcing relays or radio transmitters
10/ Bad connection on temperature sensor or faulty sensor check
connectors for continuity and mechanical soundness while watching
display for erratic behavior; check sensor and wiring for mechanical
damage.
11/ Bad connections or faulty solid state relay check connectors for
mechanical soundness and look for corrosion around terminals or signs
of arcing or other visible deterioration.
Will not maintain set point
1/ Assure that set point is at least 5 degrees over ambient.
2/ See if ambient is fluctuating; check for adjacent open doors or
HVAC duct openings stabilize ambient conditions.
Display and reference thermometer don’t
match
1/ Calibration error see section 6.4.
2/ Temperature sensor failure call Customer Service.
3/ Controller failure call Customer Service.
4/ Allow at least two hours to stabilize.
5/ Verify that reference thermometer is certified.
Can't adjust set points or calibration
1/ Turn entire unit off and on to reset.
2/ If repeatedly happens, call Customer Service.
Calibrated at one temperature, but not at
another
This can be a normal condition when operating temperature varies
widely. For maximum accuracy, calibration should be done at or as
close to the set point temperature.
MECHANICAL
Motor doesn't move; F suffix models
1/ If shaft spins freely: check connections to motor and check voltage to
motor.
2/ If shaft rubs or is frozen, relieve binding and retest.
Motor makes noise; F suffix models
1/ If noise is from the motor, tap the top of motor shaft with ball peen
hammer.
2/ If the sound gets worse, tap the other end of the shaft avoiding
touching the blower wheel.
3/ If there is no change, call Customer Service.
Door not sealing
1/ Adjust hinge blocks or twist the door.
2/ Confirm that unit has not been damaged and body is not square.
3/ Check physical condition of gasket for tears or punctures.
OTHER