Instruction Manual
23 | P a g e
OPERATION (CONTINUED)
Preparation Procedure Continued
Complete the following procedures in the Operation section:
a. Set the Temperature Set Point page 24
b. Calibrate the Temperature Display page 25
c. Set the Over Temperature Limit page 27
d. Set the CO
2
Set Point page 29
e. Calibrate the CO
2
Display page 30
f. Run a decontamination cycle page 33
End of Preparation Procedure
HUMIDIFYING THE INCUBATOR
Filling the humidity tray with distilled water and placing it on the heated incubator floor will raise the
humidity level of the chamber to approximately 90 – 95% relative humidity.
Always use the antimicrobial copper token when humidifying the incubator to slow the
growth of microbiological populations in the tray.
Regularly clean and disinfect tray.
Refill as needed, and change the water in the tray at least once per week.
Use of chemical disinfectants added to the tray may alter the surface tension of the water.
This may significantly reduce the rate of evaporation and impact the humidity level of the
incubator chamber.
Never use deionized water.
Ensure that the unit is humidified when performing temperature adjustments and
calibrations.