Operating instructions

-33
Basic Operation
The image is focused and adjusted to the screen
size using the focus ring or zoom knob on the
projector.
1 The focus is adjusted by rotat-
ing the focus ring.
2 Zooming is adjusted by moving
the zoom knob.
Focus ring Zoom knob
Zoom in
Zoom out
Downloaded From projector-manual.com Sharp Manuals
-35
Basic Operation
1 Press .
The timer starts to count down from 5 min-
utes.
On-screen display
2 Press / / / to adjust the length
of the break time.
Increases with or
5 minutes 6 minutes 60 minutes
Shortens with or
4 minutes 3 minutes 1 minute
The break time can be set in units of one
minute (up to 60 minutes).
Press .
Note
Break Timer does not function while the projec-
tor is operating the “Auto Sync”, “FREEZE” or
“AV MUTE” functions.
Press to switch the Eco+Quiet
mode between on and off.
When the Eco+Quiet Mode is set to “ON”, the sound
of the cooling fan will turn down, the power consump-
tion will decrease, and the lamp life will extend.
Note
Refer to “Checking the Lamp Life Status” on
page 53 for details of the Eco+Quiet mode.
Auto Sync function works when detecting input
signal after the projector turns on.
Press to manually adjust
with Auto Sync function.
Note
When the optimum image cannot be achieved
with Auto Sync adjustment, use manual adjust-
ments. (See page 47.)
1 Press .
The projected image is frozen.
2 Press again to return to
the moving image from the cur-
rently connected device.
MOUSE/Adjustment
buttons ( / / / )
FREEZE
button
AUTO SYNC
button
BREAK TIMER
button
ECO+QUIET
button
Canceling the break timer display function
Downloaded From projector-manual.com Sharp Manuals
-37
Basic Operation
This function allows you to modify or customize
the resize mode to enhance the input image. De-
pending on the input signal, you can choose a
desired image.
Press .
See page 49 for setting on menu screen.
COMPUTER
STRETCH
Output screen image
Input signal
Image Type
NORMAL FULL
4:3 aspect ratio
5:4 aspect ratio
15:9 aspect ratio
16:9 aspect ratio
16:10 aspect ratio
16:9 aspect ratio
Computer
Resolution
XGA and below
Resolution
higher than
XGA
SXGA
(1280 1024)
1280 720
1280 800
1360 768
1366 768
1280 768
DOT BY DOT
*1*1
*1
*3
*1
*1
*2
*2
*2
*2
*1
*1
: Cutout area on which images cannot be projected
: Area where the signals are off screen
*1 The Image Shift function can be used for these images.
*2 Same as NORMAL mode.
*3 In case SXGA+ input.
4:3 aspect ratio
Other aspect ratios
SVGA (800 600)
XGA (1024 768)
SXGA (1152 864)
SXGA+ (1400 1050)
SXGA (1280 1024)
1280 720
1360 768
1366 768
1280 768
1280 800
1068 800
1000 800
1280 720
1280 722
1280 720
1280 768
1280 800
1280 800
800 600
1024 768
1152 864
1400 1050
1280 1024
1360 768
1366 768
1280 720
1280 720
NORMAL
FULL STRETCHDOT BY DOT
Downloaded From projector-manual.com Sharp Manuals
-39
Basic Operation
When connecting the projector and the computer with a USB cable, you can use the remote control as the
computer mouse.
If the computer is placed too far away from the projector to be connected via the USB cable, the separately
sold remote receiver (AN-MR2) makes it possible to operate the projector with the remote control. For de-
tails, see the operation manual of the receiver.
Connecting with a USB cable
The mouse pointer can be operated in the
following way after it is connected.
When moving the cursor
Press MOUSE/Adjustment buttons ( / / / ).
When left-clicking
Press L-CLICK.
When right-clicking
Press R-CLICK.
When your computer supports only a
one-click mouse (such as Macintosh)
Press L-CLICK or R-CLICK.
L-CLICK and R-CLICK have common function.
Note
This function only works with the Microsoft
®
Windows
®
OS and Mac OS
®
. However, this function does not
work with the following operation systems that do not support USB.
Versions earlier than Windows
®
95
Versions earlier than Windows
®
NT4.0
Versions earlier than Mac OS
®
8.5
You cannot use this function when displaying the menu screen.
Confirm that the computer recognizes the USB connection.
When using [Page Up] or [Page Down]
Same as the [Page Up] and [Page Down] keys on a computer keyboard.
Press PAGE UP or PAGE DOWN PAGE.
To USB terminal
USB terminal
or
USB cable
(commercially available or available as Sharp service part
QCNWGA014WJPZ)
Point the remote control at
the
USB remote receiver
(AN-MR2, optional)
Computer
Downloaded From projector-manual.com Sharp Manuals
-57
Appendix
Replacing the Air Filter
¥This projector is equipped with air filters
to ensure the optimal operating condition
of the projector.
¥Ask your nearest Sharp Authorized Pro-
jector Dealer or Service Center to ex-
change the filter (PFILDA022WJKZ) when
it is no longer possible to clean.
1 Press on the pro-
jector or on the remote
control to put the projector into the
standby mode.
Wait until the cooling fan stops.
2 Disconnect the power cord.
Unplug the power cord from the AC socket.
3 Remove the filter cover.
Turn over the projector. Pull down the car-
rying handle ( ), grab the tab ( ) and slide
open the filter cover ( ).
4 Remove the air filter.
Pick the air filter up with your fingers and lift
it out of the filter cover.
STANDBY button
AC socket
Tab
2
1
3
STANDBY/ON
button
Downloaded From projector-manual.com Sharp Manuals
-59
Appendix
Power indicator
Lamp indicator
Temperature warning indicator
Maintenance Indicators
Maintenance Indicators
The warning lights on the projector indicate problems inside the projector.
If a problem occurs, either the temperature indicator or the lamp indicator will illuminate red, and
the projector will enter the standby mode. After the projector has entered the standby mode, follow
the procedures given below.
About the temperature warning indicator
If the temperature inside the projector increases, due to blockage of the air vents, or the setting location,
” will illuminate in the lower left corner of the picture. If the temperature keeps on rising, the lamp will
turn off and the temperature warning indicator will blink, the cooling fan will run for a further 90 seconds, and
then the projector will enter the standby mode. After “
appears, ensure you perform the measures
described on page 60.
About the lamp indicator
When the remaining lamp life becomes 5% or less, (yellow) and “Change The Lamp. will be
displayed on the screen. When the percentage becomes 0%, it will change to “
(red), the lamp will
automatically turn off and then the projector will automatically enter the standby mode. At this time,
the lamp indicator will illuminate in red.
If you try to turn on the projector a fourth time without replacing the lamp, the projector will not turn on.
Downloaded From projector-manual.com Sharp Manuals
-61
Appendix
It is recommended that the lamp unit (optional: AN-C430LP) be replaced when the remaining lamp
life becomes 5% or less, or when you notice a significant deterioration in the picture and color
quality. The lamp life (percentage) can be checked with the on-screen display. See page 53.
Purchase a replacement lamp unit of type AN-C430LP from your place of purchase, nearest Sharp
Authorized Projector Dealer or Service Center.
IMPORTANT NOTE TO U.S. CUSTOMERS:
The lamp included with this projector is backed by a 90-day parts and labor limited warranty. All service
of this projector under warranty, including lamp replacement, must be obtained through a Sharp Autho-
rized Projector Dealer or Service Center. For the name of the nearest Sharp Authorized Projector Dealer
or Service Center, please call toll-free: 1-888-GO-SHARP (1-888-467-4277).
Hg LAMP CONTAINS MERCURY For State Lamp Disposal Information
www.lamprecycle.org or 1-800-BE-SHARP
This projector utilizes a pressurized mercury lamp. A loud sound may indicate lamp failure. Lamp
failure can be attributed to numerous sources such as: excessive shock, improper cooling, surface
scratches or deterioration of the lamp due to a lapse of usage time.
The period of time up to failure largely varies depending on the individual lamp and/or the condition
and the frequency of use. It is important to note that failure can often result in the bulb cracking.
When the lamp replacement indicator and on-screen display icon are illuminated, it is recommended
that the lamp be replaced with a new one immediately, even if the lamp appears to be operating
normally.
Should the lamp break, the glass particles may spread inside the lamp cage or gas contained in the
lamp may be vented into the room from the exhaust vent. Because the gas in this lamp includes
mercury, ventilate the room well if the lamp breaks and avoid all exposure to the released gas. In
case of exposure to the gas, consult a doctor as soon as possible.
Should the lamp break, there is also a possibility that glass particles may spread inside of the
projector. In such a case, it is recommended you contact your nearest Sharp Authorized Projector
Dealer or Service Center to assure safe operation.
Caution
Do not remove the lamp unit from the projector right after use. The lamp will be very hot and may cause burn
or injury.
Wait at least one hour after the power cord is disconnected to allow the surface of the lamp unit to fully cool
before removing the lamp unit.
Carefully change the lamp by following the instructions described in this section. *If you wish, you
may have the lamp replaced at your nearest Sharp Authorized Projector Dealer or Service Center.
* If the new lamp does not illuminate after replacement, take your projector to the nearest Sharp Authorized
Projector Dealer or Service Center for repair.
Regarding the Lamp
U.S.A. ONLY
Downloaded From projector-manual.com Sharp Manuals