User guide
TM200 V4.1 11/06 6
Specifications
Measurement Range -328 to 2501
°
F (-200 to 1372
°
C)*
Max. Resolution 0.1
°
C/
°
F<999.9
°
, 1
°
C/
°
F>1000
o
-148 to 2501
°
F (-100 to 1372
°
C) ± [0.5% of reading + 1.8
o
F(1
°
C)] Accuracy (T1&T2)
-328 to -148
°
F (-200 to -100
°
C) ± [0.5% of reading + 3.6
o
F(2
°
C)]
Temperature
Coefficient
0.1 times the applicable accuracy specifications per
°
C from 0
°
C to
18
°
C and 28
°
C to 50
°
C (32
°
F to 64
°
F and 82
°
F to 122
°
F)
Note: Temperature accuracy does not include the accuracy of the type K probe
Note: Temperature accuracy specifies @ 23 ± 5
o
C
Note: The temperature scale is based on the international temperature scale of 1990
(ITS90)
Note: Supplied thermocouples rated for 500
°
F (260
°
C) max.
General Specifications
Display Multi-function LCD with Backlight
Input Protection 60VDC; 24VAC rms
Display Update Rate 2.5 times per second
Over range indication "OL" appears on the LCD
Open input indication Dashes "-------" appears on the LCD
Low battery indication “ ” appears on the LCD
Power supply 9V alkaline battery
Auto Power OFF After 20 minutes of inactivity (programmable)
Ambient Operating conditions 4 to 122
o
F (0 to 50
o
C) / < 80% Relative Humidity
Ambient Storage conditions 14 to 122
o
F (-10 to 50
o
C) / 10 to 80% Relative Humidity
Dimensions 6.5 x 3.0 x 1.7" (165 x 76 x 43mm) without holster
Weight 7.4oz (210g) with battery