Instruction manual

Operation
Volumetric Conditions
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-25
The Pressure Compensation screen is used to turn the pressure correction
to actual or standard.
Press to toggle and set the pressure compensation to actual or
standard.
PRESSURE:
>COMPENSATION 760 mmHg
STANDARD 760 mmHg
RANGE AVG DIAGS ALARM
The Standard Pressure screen is used to set the standard pressure.
In the Main Menu, choose Instrument Controls > Volumetric
Conditions > Pressure > Standard.
STANDARD PRESSURE:
CURRENTLY: ACT
SET TO: 760 mmHg ?
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The iSeries instruments include a built-in datalogging capability as a
standard feature. The operator is allowed to create two different types of
records, which for historical reasons, are named lrecs and srecs. Each record
can contain up to 32 different fields, or data items, and records can be
created at user-defined intervals ranging from 1 to 60 minutes.
Record generation is tied to the instrument’s real-time clock. For example,
if the logging period for srecs is set to 30 minutes, a new srec will be
generated on every hour and every half hour (10:00, 10:30, 11:00 …).
Lrecs and srecs function independent of each other. So, for example, an
srec containing just the current SHARP level could be generated every 5
minutes while an lrec containing a full set of diagnostic data could be
generated once every hour.
The analyzer's computer system includes three megabytes of flash memory
which is enough to store a full lrec containing 32 data items and a full srec
Pressure Compensation
Standard Pressure
Datalogging
Settings