Instruction manual

Operation
Volumetric Conditions
3-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
The Volumetric Conditions menu allows the user to turn
temperature/pressure compensation on and off and to set the standard
pressure and temperature values in various modes.
In the Main Menu, choose Instrument Controls > Volumetric
Conditions.
VOLUMETRIC CONDITIONS:
>TEMPERATURE ACT
PRESSURE STD
RANGE AVG DIAGS ALARM
The Temperature screen is used to turn the temperature correction to
actual or standard.
Press to toggle and set the temperature compensation to actual
or standard.
TEMPERATURE:
>COMPENSATION ACT
STANDARD 25
o
C
RANGE AVG DIAGS ALARM
The Standard Temperature screen is used to set the standard temperature.
In the Main Menu, choose Instrument Controls > Volumetric
Conditions > Temperature > Standard.
STANDARD TEMPERATURE:
CURRENTLY: 25
o
C
SET TO: 0.0
o
C ?
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Volumetric
Conditions
Temperature
Compensation
Standard Temperature