Instruction manual

Operation
Instrument Controls Menu
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-23
In the Main Menu, choose Instrument Controls > Filter Tape
Control > Period.
PERIOD:
CURRENTLY: 24 HRS
SET TO: 000008 HRS?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Counter screen allows the user to record the current filter tape count
and reset the tape count to zero.
In the Main Menu, choose Instrument Controls > Filter Tape
Control > Counter.
FILTER TAPE COUNTER: 0
RESET TO ZERO
RANGE AVG DIAGS ALARM
The Tape/Zero Ratio screen allows the user to set the occurrence of
nephelometer zeroing after X filter changes. For PM
2.5
X = 42 and for PM
10
X = 14 based on a 2-week zeroing cycle. A conservative ratio of 1:1 will
zero the nephelometer with every filter tape change.
In the Main Menu, choose Instrument Controls > Filter Tape
Control > Tape/Zero Ratio.
TAPE/NEPH ZERO RATIO:
CURRENTLY: 1:1
SET TO: 2:1 ?
©ª CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Counter
Tape/Zero Ratio