Instruction manual

Operation
Instrument Controls Menu
3-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
RESTORE PREV BKG VALUES?
RESTORE
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
The Nephelometer Coefficient screen allows the user to manually adjust
the Nephelometer coefficient. The Nephelometer concentration based on
the SET COEF TO is displayed to facilitate setting. The Nephelometer
concentration displayed in the Run portion of the screen is NOT affected
by the SET COEF until the value is saved.
In the Main Menu, choose Calibration Factors > Neph Coef.
NEPH COEFFICIENT:
CURRENTLY: 227.346
SET TO: 232.500 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Instrument Controls menu contains a number of items that may be
selected to control various instrument operational parameters. The
firmware controls listed in this menu enable control of the listed
instrument functions.
In the Main Menu, choose Instrument Controls.
Nephelometer
Coefficient
Instrument Controls
Menu
RESTORE PREV BKG VALUES?
RESTORE
RANGE AVG DIAGS ALARM