Instruction manual

Operation
Integration Time
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-13
In the Main Menu, choose Range > Set Custom Ranges.
CUSTOM RANGES:
>CUSTOM RANGE 1 555.6
CUSTOM RANGE 2 1875.0
CUSTOM RANGE 3 8125.0
RANGE AVG DIAGS ALARM
The Custom Ranges screen is used to define the custom ranges.
The display shows the current custom range. The next line of the display is
used to set the range. For more information about selecting ranges, see
SHARP Range” above.
In the Main Menu, choose Range > Set Custom Ranges > Custom
Range 1, 2, or 3.
CUSTOM RANGE 1:
CURRENTLY: 555.6
SET TO: 000555.7 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Integration Time screen allows the user to adjust the compounded
averaging of the beta attenuation mass and concentration. The default
value should be set to 20 minutes and may be adjusted between 15–60
minutes. The display shows the current integration time is set to 15. The
next line of the display is used to change the integration time.
In the Main Menu, choose Integration Time.
INTEGRATION TIME:
CURRENTLY: 15 MIN
SET TO: 20 MIN?
©ª CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Custom Ranges
Integration Time