Instruction manual

Operation
Range Menu
3-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
XXLO WCONCENTRATIONXXLO W
SHARP 15.0
3µg/m3
x 12:34 x
MAIN MENU:
>RANGE
INTEGRATION TIME
24HR AVERAGE
CALIBRATION FACTORS
INSTRUMENT CONTROLS
DIAGNOSTICS
ALARMS
ª
RANGE AVG DIAGS ALARM
SERVICE
PASSWORD
The Range menu allows the operator to select the concentration units, PM
standard ranges, and to set the custom ranges.
In the Main Menu, choose Range
RANGE:
>CONC UNITS 3μg/m3
XSHARP RANGE 10000
SET CUSTOM RANGES
RANGE AVG DIAGS ALARM
The Concentration Units screen defines how the particles concentration
readings are expressed. Concentration units of micrograms per cubic meter
(μg/m
3
) and milligrams per cubic meter (mg/m
3
) are available.
When switching the selected units, the instrument’s analog ranges all
default to the highest range. For example, when switching from mg/m
3
to
μg/m
3
, the ranges default to 10000 μg/m
3
. Therefore, whenever you change
units, you should also check the range settings.
Note If the units change, the instrument should be re-calibrated. A display
warning will appear.
In the Main Menu, choose Range > Conc Units.
Range Menu
Concentration Units