Instruction manual
Installation
Acceptance Testing and Startup Procedures
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-7
Figure 2–2. Model 5030i Sampling Room Air Under Stable Conditions
Record the temperature sensor readings from the Model 5030i and
compare to your NIST-traceable thermometer. Each of these measurements
should be within ±2 °C tolerance of your NIST-traceable thermometer.
Within this tolerance, the temperature sensors have passed the acceptance
test.
If the sensors are slightly out of tolerance (±3 °C), the acceptance test
should be classified as marginal. If the sensor performance is outside this
range, please perform a temperature sensor calibration.
For more information about the temperature sensor calibration, see the
“Calibration” chapter.
Record the ambient RH sensor and nephelometer RH sensor readings from
the Model 5030i and compare to your NIST-traceable Hygrometer. The
Model 5030i RH sensors should compare within ±2% RH tolerance of
your NIST-traceable hygrometer. If the RH sensor performance is within
this tolerance, the acceptance test has passed.
Note The ambient temperature verification should be completed prior to
performing the RH-sensor verification due to a thermal compensation
applied to the RH-sensor. Furthermore, try to avoid RH comparisons
<30% RH and >80% RH. ▲