Thermo Fisher Scientific Model 5030i SHARP Instruction Manual E-1
Appendix E
ESM Protocol Commands
This appendix provides a description of the ESM Protocol commands
(from the prior FH62 platform) that can be used to remotely control a
Model 5030i instrument using a host device such as a PC or a datalogger.
ESM protocol may be used over RS-232, RS-485 or over Ethernet. ESM
functions can be accessed over Ethernet using TCP port 9884.
Up to three simultaneous connections per protocol may be made over
Ethernet.
Table E–1 through Table E–3 lists the ESM commands supported for the
Model 5030i.
Table E–1. Read Commands for 5030i
Command Description
C3 PM conc always always µg/m³
C2 Neph conc µg/m³
C SHARP conc µg/m³
H3 Avg PM conc always µg/m³
H2 Avg Neph conc µg/m³
HT Avg SHARP conc µg/m³
JB temperature sampling head–ambient temp
JD temperature inside of the flow meter orifice–flow temp
JF different pressure of the air flow measuring module in Pa
JG low pressure suction chamber
JH barometer in hPa
JI air flow sampling head–volumetric flow
JJ norm air flow (273K, 1013hPa)
JR relative humidity value (%RH)
JS sample relative humidity value (%RH)
m1 non filtered mass in µg or mg based on units selected
o reports “offline” in service mode, “online” if NOT in service mode
ESM Commands
Supported