Instruction manual

Gesytec (Bayern-Hessen) Protocol
Gesytec Commands
D-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
See Table D–1 for operating status and Table D–2 for error status for the
Model 5030i.
Table D–1. Operating Status for Model 5030i
D7 D6 D5 D4 D3 D2 D1 D0
Æ Bit 8 7 6 5 4 3 2 1
Æ Hex-value 80 40 20 10 08 04 02 01
MSB LSB
Operating status:
Service Mode (On) 0 0 0 0 0 0 0 1
No Password Protect 0 0 0 0 0 0 1 0
Pump (Off) 0 0 0 0 0 1 0 0
Heater (Off) 0 0 0 0 1 0 0 0
Filter Change 0 0 0 1 0 0 0 0
Nephelometer Zero 0 0 1 0 0 0 0 0
Nephelometer Standby 0 1 0 0 0 0 0 0
Not Used 1 0 0 0 0 0 0 0
Table D–2. Error Status for Model 5030i
D7 D6 D5 D4 D3 D2 D1 D0
Æ Bit 8 7 6 5 4 3 2 1
Æ Hex-value 80 40 20 10 08 04 02 01
MSB LSB
Error status:
Filter Tape Cntr Alarm 0 0 0 0 0 0 0 1
Filter Change 0 0 0 0 0 0 1 0
Beta Cnts Alarm 0 0 0 0 0 1 0 0
Neph Brd Alarm 0 0 0 0 1 0 0 0
Detector Board Status Alarm 0 0 0 1 0 0 0 0
Motherboard Status Alarm 0 0 1 0 0 0 0 0
Flow Alarm 0 1 0 0 0 0 0 0
Nephelometer Reference Voltage
Alarm
1 0 0 0 0 0 0 0
Operating and Error
Status