Instruction manual

Gesytec (Bayern-Hessen) Protocol
Gesytec Commands
D-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
The Gesytec serial number defaults to zero. To set the Gesytec serial
number select Main Menu > Instrument Controls > Communication
Settings > Gesytec Serial No.
Example of response to DA command from an instrument with Gesytec
serial number set to 000. The Gesytec serial number is bold in the
example.
Gesytec Protocol with transmission of three concentrations (Instrument ID
is 1, Operation Status is 03, Error Status is 04):
Data Query String: <STX>DA<CR>
Reply String:
<STX>MD03<SP>001<SP>+2578+01<SP>03 <SP>04<SP>0000000000 <SP>002 <SP>
Address First Concentration(E-format)=25.78 Address+1
+5681+00<SP>03<SP>04<SP>0000000000<SP>003<SP>+1175+01<SP>03<SP>04<SP
Second Concentration = 5.681 Address+2 Third Concentration=11.75
0000000000<SP><CR>
Example of response to DA command from an instrument with Gesytec
serial number set to 123. The Gesytec serial number is bold in the
example.
Gesytec Protocol with transmission of three concentrations (Instrument ID
is 1, Operation Status is 03, Error Status is 04):
Data Query String: <STX>DA<CR>
Reply String:
<STX>MD03<SP>001<SP>+2578+01<SP>03 <SP>04<SP>1230000000 <SP>002 <SP>
Address First Concentration(E-format)=25.78 Address+1
+5681+00<SP>03<SP>04<SP>0000000000<SP>003<SP>+1175+01<SP>03<SP>04<SP
Second Concentration = 5.681 Address+2 Third Concentration=11.75
1230000000<SP><CR>
The attached concentrations are in the selected gas units. The
measurements that are attached to the response, if not valid in a particular
mode, will report a value of 0.0.