Instruction manual

MODBUS Protocol
MODBUS Parameters Supported
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-11
Register Number Variable
91 & 92 EXT ALARMS
93 & 94 NEPH TEMP
95 & 96 NEPH RH
97 & 98 NEPH IRED
99 & 100 NEPH REF
101 & 102 BETA REF
103 & 104 NOT USED
105 & 106 COMMON FLAGS
107 & 108 BETA DETECTOR FLAGS
109 & 110 NOT USED
111 & 112 COMP MASS
113 & 114 NOT USED
115 & 116 AVG NEPH
IMPORTANT NOTE The addresses in the following tables are Protocol
Data Unit (PDU) addresses. Verify the coil number on your MODBUS
master to ensure that it matches the coil number on the instrument.
Note Writing 1 to the coil number shown in the following table will
initiate the “action triggered” listed in the table. This state must be held for
at least 1 second to ensure the instrument detects the change and triggers
the appropriate action.
Note The coils within each coil group in the following table are mutually
exclusive and will not be triggered if there is conflict. Before you assert (1)
one coil in a group, make sure the other coils in the group are de-asserted
(0).
Table C–3. Write Coils for 5030i
Coil Number Action Triggered
100 INVALID
101 NOT USED
102 NOT USED
103 NOT USED
104 AOUTS TO ZERO
105 AOUTS TO FS