Instruction manual
C-Link Protocol Commands
I/O Configuration
B-64 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
D to A Function Range
8 Current Output AVG PM
9 Current Output Flow
10 Current Output Ambient Temperature
11 Current Output Flow Temperature
12 Current Output Ambient Relative Humidity
list din
list dout
These commands report the current selection for the digital inputs or the
digital outputs in the format. Output no Index number variable name
active state. The active state for digital outputs is open or closed. The active
state for digital inputs is high or low.
Send: list dout
Receive: list dout
output index variable state
1 5 CONC ALARM open
2 1 LOCAL/REMOTE open
3 3 UNITS open
list var aout
list var dout
list var din
list var log
These commands report the list of possible index numbers, and the
variables (associated with that index number) available for selection in the
current mode for analog outputs, digital outputs, digital inputs and
datalogging. The index number is used to insert the variable in a field
location in a list using “set sp field index”. The following example reports
the list of analog outputs, index numbers, and variables.
Send: list var aout
Receive: list var aout
index variable
0 none
1 pm
3 baro
4 vac
6 pflow
8 ambrt
9 srh
11 ambtemp
12 brdtemp
13 ftemp
15 stemp