C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-63
dout channel
This command reports the index number, output variable and the active
state assigned to digital output channel. The following example reports
output 4 to be assigned an index number 11 corresponding to action of
“flowl alarms”.
Send: dout 4
Receive: dout 4 11 FLOW ALARMS open
set dout channel index state
This command assigns digital output channel to be assigned to the action
associated with index, and assigns it an active state of state (open or closed).
Use the “list var dout” command to obtain the list of supported index
values and corresponding state. The following example sets the digital
output channel 4 to state 11.
Send: set dout 4 11 open
Receive: set dout 4 11 open ok
dtoa channel
This command reports the outputs of the 6 or 12 Digital to Analog
converters, according to Table B–13. The following example shows that the
D/A #1 is 97.7% full-scale.
Send: dtoa 1
Receive: dtoa 1 97.7%
Note If the instrument is in a mode which does not provide a particular
output, and that output is selected, the value will be 0.0.
All channel ranges are user definable. If any customization has been made
to the analog output configuration, the default selections may not apply.
Table B–13. Default Analog Output Channel Assignments
D to A Function Range
1 Voltage Output PM
2 Voltage Output AVG PM
3 Voltage Output Flow
4 Voltage Output Ambient Temperature
5 Voltage Output Flow Temperature
6 Voltage Output Ambient Relative Humidity
7 Current Output PM