Instruction manual

C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-61
analog iout range channel
This command reports the analog current output range setting for channel,
where channel must be between 1 and 6, inclusive. The following example
reports current output channel 4 is in the 4-20 mA range, according to
Table B–11. This command responds with “feature not enabled” if the I/O
expansion board is not detected.
Send: analog iout range 4
Receive: analog iout range 4 2
set analog iout range channel range
This command sets analog current output channel to the range where
channel is between 1 and 6 inclusive, and range is set according to Table B–
11. The following example sets current output channel 4 to the 0-20 mA
range. This command responds with “feature not enabled” if the I/O
expansion board is not detected.
Send: set analog iout range 4 1
Receive: set analog iout range 4 1 ok
Table B–11. Analog Current Output Range Values
Range Output Range
1 0-20 mA
2 4-20 mA
0 [cannot be set to this, but may report] Undefined
analog vin channel
This command retrieves the analog voltage input channel data, both the
calculated value and the actual voltage. In the following example, the
“calculated” value of channel 1 is 75.325 degrees F, volts are 2.796. This
command responds with “feature not enabled” if the I/O expansion board
is not detected.
Send: analog vin 1
Receive: analog vin 1 75.325, 2.796 V
analog vout range channel
This command reports the analog voltage output channel range, where
channel is between 1 and 6 inclusive, according to Table B–12. The
following example reports that analog voltage output channel 2 is set to 3
(0-10 V).
I/O Configuration