Instruction manual

C-Link Protocol Commands
Communications Configuration
B-58 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
set format format
This command sets the reply termination format, where format is set
according to Table B–8. The following example sets the reply termination
format to checksum.
Send: set format 01
Receive: set format 01 ok
Table B–8. Reply Termination Formats
Format Reply Termination
00 <CR>
01 <NL> sum xxxx <CR>
where xxxx = 4 hexadecimal digits that represent the sum of all the
characters (bytes) in the message
host name
This command reports the host name string. The following example
reports the host name is set to iSeries.
Send: host name
Receive: host name iSeries
set host name string
This command sets the host name string, where string is 1-13 alphanumeric
characters. The following example sets the host name to analyzer01.
Send: set host name analyzer01
Receive: set host name analyzer01 ok
instr name
This command reports the instrument name. The following example
reports the instrument name for the Model 5030i.
Send: instr name
Receive: instr name
SHARP
Particle Analyzer
instrument id
This command reports the instrument identification (ID). The following
example reports the current setting of the instrument ID.
Send: instrument id
Receive: instrument id 14