Instruction manual
C-Link Protocol Commands
Hardware Configuration
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-51
set temp std value
This command sets the standard temperature to value, where value is a
floating-point number representation of standard temperature between 0 to
25 °C. The following example sets the standard temperature to 15 °C.
Send: set temp std
Receive: set temp std 15 ok
temp th
This command reports the heater temperature threshold for the sample.
The following example reports that the temperature threshold is 30 °C.
Send: temp th
Receive: temp th 30 degC
set temp th value
This command sets the heater temperature threshold for the sample to
value, where value is a floating-point number representation of temperature
threshold between 0 to 70 °C. The following example sets the temperature
threshold to 35 °C.
Send: set temp th 35
Receive: set temp th 35 ok
contrast
This command reports the screen’s level of contrast. The following example
shows the screen contrast is 55%, according to Table B–6.
Send: contrast
Receive: contrast 11: 55%
set contrast level
This command sets the screen’s level of contrast, according to Table B–6.
The following example sets the contrast level to 50%.
Send: set contrast 10
Receive: set contrast 10 ok
Table B–6. Contrast Levels
Level Contrast Level
0 0%
1 5%
2 10%
3 15%
4 20%