Instruction manual
C-Link Protocol Commands
Measurement Configuration
B-50 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Selection μg/m
3
mg/m
3
6 C1 C1
7 C2 C2
8 C3 C3
rh th
This command reports the current heater RH threshold for the sample.
The following example reports that the RH threshold is 30%.
Send: rh th
Receive: rh th 30 %
set rh th value
This command sets the heater RH threshold for the sample to value, where
value is a floating-point number representation of RH threshold between
25 to 90%. The following example sets the RH threshold to 35%.
Send: set rh th 35
Receive: set rh th 35 ok
temp comp
This command reports whether temperature compensation is for actual or
standard conditions. The following example shows the temperature
compensation is standard.
Send: temp comp
Receive: temp comp std
set temp comp selection
selection = | act | std |
These commands turn the temperature compensation to either actual or
standard conditions. The following example turns the temperature
compensation to actual.
Send: set temp comp act
Receive: set temp comp act ok
temp std
This command reports the standard temperature. The following example
reports that the standard temperature is 18 °C.
Send: temp std
Receive: temp std 18 degC