C-Link Protocol Commands
Measurement Configuration
B-48 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
This command sets the heater to off, RH threshold, or temp threshold.
The following example sets the heater to temp threshold.
Send: set ht control temp
Receive: set ht control temp ok
int time
This command reports the current integration time of the calculated mass
and concentration. The following example reports the integration time is
15 minutes.
Send: int time
Receive: int time 15 min
set int time value
This command sets the integration time to value, where value can be 15,
20, 30, 40, 45, or 60 minutes. The following example sets the integration
time to 20 minutes.
Send: set int time 20
Receive: set int time 20 ok
mass limit
This command reports the current filter mass limit. The following example
reports that the mass limit is 1,500 mg.
Send: mass limit
Receive: mass limit 1500
set mass limit value
This command sets the filter mass limit to value, where value is a floating-
point number representation of mass limit between 0.5 to 5.0 mg or 500 to
5,000 μg. The following example sets the mass limit to 1.0 mg/m
3
.
Send: set mass limit 1.0
Receive: set mass limit 1.0 ok
pres comp
This command reports whether pressure compensation is for actual or
standard conditions. The following example shows the pressure
compensation is standard.
Send: pres comp
Receive: pres comp std
set pres comp selection
selection = | act | std |