Instruction manual

C-Link Protocol Commands
Measurement Configuration
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-47
filter time
This command reports the next date and time of the filter tape change. The
following example reports that the date and time is January 1, 2008, 6:12
pm.
Send: filter time
Receive: filter time 01Jan08 18:12
set filter time dd-mm-yyyy hh:mm:ss
dd = day
mm = month
yyyy = year
hh = hours
mm = minutes
ss = seconds
This command sets the next filter tape change. The following example sets
the filter time to October 1, 2008, 11:32 pm.
Send: set filter time 10-01-2008 23:32
Receive: set filter time 10-01-2008 23:32 ok
flow target
This command reports the current flow. The following example reports
that the flow is 16.0 LPM.
Send: flow target
Receive: flow target 16.00
set flow target value
This command sets the flow to value, where value is a floating-point
number representation of flow between 1 to 20 LPM. The following
example sets the flow to 16.67 LPM.
Send: set flow target 16.67
Receive: set flow target 16.67 ok
ht control
This command reports the current status of the heater (off, RH threshold,
or temp threshold). The following example reports that the heater is RH
threshold.
Send: ht control
Receive: ht control RH
set ht control selection
selection = | off | rh | temp |