Instruction manual
C-Link Protocol Commands
Measurement Configuration
B-46 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Send: conc unit
Receive: conc unit μg/m3
set conc unit unit
unit = | μg/m
3
| mg/m
3
|
This command sets the concentration units to mg/m
3
or μg/m
3
. The
following example sets the concentration units to mg/m
3
.
Send: set conc unit mg/m3
Receive: set conc unit mg/m3 ok
custom range
range = | 1 | 2 | 3 |
This command reports the user-defined value of custom range 1, 2, or 3.
The following example reports that custom range 1 is defined to 5.50
mg/m
3
.
Send: custom 1
Receive: custom 1 5.500E+00 mg/m3
set custom range range value
set custom 1 value
set custom 2 value
set custom 3 value
These commands are used to set the maximum concentration for any of the
three custom ranges 1, 2, or 3 to range value, where value is a floating-point
number representing concentration in the currently selected units. The
following example sets the custom 1 range to 100.5 μg/m
3
.
Send: set custom 1 100.5 μg/m3
Receive: set custom 1 100.5 μg/m3 ok
filter period
This command reports the number of hours to be added to filter next time
for the filter tape change. The following example reports that the filter
period is eight hours.
Send: filter period
Receive: filter period 8 HRS
set filter period value
This command sets the number of hours to be added to filter next time for
the filter tape change to value, where value is a floating-point representation
of the filter period in hours. The following example sets the filter period to
10 hours.
Send: set filter period 10
Receive: set filter period 10 ok