Instruction manual

C-Link Protocol Commands
Keys/Display
B-44 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
set vac pres span value
This command sets the vacuum span to the value specified at the time of
calibration, where value is a floating-point representation of vacuum span.
The following example successfully sets the new value for vacuum pressure
span to 1.002.
Send: set vac pres span 1.002
Receive: set vac pres span 1.002 ok
vac pres offset
This command reports the vacuum offset for the detector at the time of
calibration. The following example reports that the vacuum pressure offset
is 0.3.
Send: vac pres zero
Receive: vac pres zero 0.3
set vf pres cal def
This command resets the vacuum and flow pressure calibration to their
default values.
Send: set vf pres cal def
Receive: set vf pres cal def ok
push button
do me 1
down menu 2
en ri 3
enter right 4
he ru
help run
le up
left
button = | do | down | en | enter | he | help | le | left | me | menu | ri | right |
ru | run | up | 1 | 2 | 3 | 4 |
These commands simulate pressing the front panel pushbuttons. The
numbers represent the front panel soft keys, from left to right.
Send: push enter
Receive: push enter ok
isc
iscreen
These commands retrieve the framebuffer data used for the display on the
iSeries instrument. It is 19200 bytes in size, 2-bits per pixel, 4 pixels per
byte arranged as 320 by 240 characters. The data is sent in RLE encoded
Keys/Display