Instruction manual

C-Link Protocol Commands
Calibration
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-43
set neph temp offset value
This command sets the nephelometer temperature offset to user-defined
values to value, where value is a floating-point representation of the offset.
The following example sets the nephelometer temperature offset to 2.5 °C.
Send: set neph temp offset 2.5
Receive: set neph temp offset 2.5 ok
neph thermal coef
This command reports the nephelometer thermal coefficient set at the
factory. The following example reports that the nephelometer thermal
coefficient is 0.00058.
Send: neph thermal coef
Receive: neph thermal coef 0.00058
set neph thermal coef value
This command sets the nephelometer thermal coefficient to user-defined
values to value, where value is a floating-point representation of the
coefficient. The following example sets the nephelometer thermal
coefficient to 0.001.
Send: set neph thermal coef 0.001
Receive: set neph thermal coef 0.001 ok
tape to ncal ratio
This command reports the ratio of filter tape changes per nephelometer
auto-zero. The following example reports this value.
Send: tape to ncal ratio
Receive: tape to ncal ratio 1:1
set tape to ncal ratio
This command sets the filter tape change per nephelometer auto-zero ratio
to a user-defined value, where value is a floating-point representation of the
ratio. The following example sets the nephelometer filter tape change to
auto-zero to 2:1.
Send: set tape to ncal ratio 2:1
Receive: set tape to ncal ratio 2:1 ok
vac pres span
This command reports the vacuum span for the detector at the time of
calibration. The following example reports that the vacuum pressure span is
1.0.
Send: vac pres span
Receive: vac pres span 1.000