Instruction manual
C-Link Protocol Commands
Calibration
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-41
Send: set thermal mass coef 20.5
Receive: set thermal mass coef 20.5 ok
vac mass coef
This command reports the current vacuum mass coefficient. The following
example reports that the vacuum mass coefficient is 2.0.
Send: vac mass coef
Receive: vac mass coef 2.000000
set vac mass coef value
This command sets the vacuum mass coefficient to user-defined values to
value, where value is a floating-point representation of the coefficient. The
following example sets the vacuum mass coefficient to 3.1.
Send: set vac mass coef 3.1
Receive: set vac mass coef 3.1 ok
neph cal bkg
This command sets the nephelometer into an auto-zero calibration routine.
The following example set this command.
Send: set neph cal bkg
Receive: set neph cal bkg ok
neph coef
This command reports the nephelometer span coefficient set at the factory.
The following example reports that the nephelometer span coefficient is
388.000.
Send: neph coef
Receive: neph coef 388.000
set neph coef value
This command sets the nephelometer span coefficient to user-defined
values to value, where value is a floating-point representation of the
coefficient. The following example sets the nephelometer span coefficient
to 250.00
Send: set neph coef 250
Receive: set neph coef 250 ok
neph reset bkg
This command will reset all measured background values to zero. The
following example sets this command.
Send: set neph reset bkg
Receive: set neph reset bkg ok