Instruction manual
C-Link Protocol Commands
Calibration
B-40 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
set flow temp offset value
This command sets the offset value for flow temperature, where value is a
floating-point number representing flow temperature offset in degrees C.
The following example sets the flow temperature offset to 7 °C.
Send: set flow temp offset 7
Receive: set flow temp offset 7 ok
high volt
This command reports the current high voltage for the detector. The
following example reports that the high voltage is 1400 V.
Send: high volt
Receive: high volt 1400
set high volt value
This command sets the high voltage for the detector to value, where value is
a floating-point representation of the high voltage. The following example
sets the high voltage to 1445 V.
Send: set high volt 1445
Receive: set high volt 1445 ok
mass coef
This command reports the current mass coefficient. The following example
reports that the mass coefficient is 7100.
Send: mass coef
Receive: mass coef 7100.0
set mass coef value
This command sets the mass coefficient to user-defined values to value,
where value is a floating-point representation of the coefficient. The
following example sets the mass coefficient to 7000.
Send: set mass coef 7000
Receive: set mass coef 7000 ok
thermal mass coef
This command reports the current thermal mass coefficient. The following
example reports that the thermal mass coefficient is 19.000000.
Send: thermal mass coef
Receive: thermal mass coef 19.000000
set thermal mass coef value
This command sets the thermal mass coefficient to user-defined values to
value, where value is a floating-point representation of the coefficient. The
following example sets the thermal mass coefficient to 20.5.