C-Link Protocol Commands
Calibration
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-39
flow pres span
This command reports the flow pressure span recorded at the time of
calibration. The following example reports that the flow span is 1.2.
Send: flow pres span
Receive: flow pres span 1.200
set flow pres span value
This command sets the flow pressure span to value, where value is a
floating-point representation of the pressure span specified at the time of
calibration. The following example successfully sets the new value for flow
pressure to 1.5.
Send: set flow pres span 1.5
Receive: set flow pres span 1.5 ok
flow pres offset
This command reports the flow pressure offset for the detector at the time
of calibration. The following example reports that the flow pressure offset is
3.0.
Send: flow pres offset
Receive: flow pres offset 3.0
flow span
This command reports the current flow span at the time of calibration. The
following example reports that the flow at calibration is 1.00.
Send: flow span
Receive: flow span 1.000
set flow span value
This command sets the current flow span to the value, where value is a
floating-point representation of the flow span specified at the time of
calibration. The following example sets the flow span to 1.5.
Send: set flow span 1.5
Receive: set flow span 1.5 ok
flow temp offset
This command reports offset for flow temperature, in degrees C. The
following example reports that the flow temperature offset is 5 °C.
Send: flow temp offset
Receive: flow temp offset 5.0 degC