Instruction manual

C-Link Protocol Commands
Calibration
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-37
set bkg pm value
This command is used to set PM backgrounds to user-defined values to
value, where value is a floating-point representation of the background in
current selected units. The following example sets the PM background to
5.2 μg/m
3
.
Send: set bkg pm 5.2
Receive: set bkg pm 5.2 ok
bkg sharp
This command reports the current SHARP backgrounds. The following
example reports that the SHARP background is 4 μg/m
3
.
Send: bkg sharp
Receive: bkg sharp 4.000E+00 μg/m3
set bkg sharp value
This command is used to set SHARP backgrounds to user-defined values to
value, where value is a floating-point representation of the background in
current selected units. The following example sets the SHARP background
to 5.2 μg/m
3
.
Send: set bkg sharp 5.2
Receive: set bkg sharp 5.2 ok
set cal baro pres span
This command automatically calculates the span coefficient based on the
entered target pressure value. The following example sets the calibration
pressure to 720.5 mmHg.
Send: set cal baro pres span 720.5
Receive: set cal baro pres span 720.5 ok
set cal flow pres span value
This command automatically sets the flow pressure span based on the flow
pressure entered at the time of calibration. The following example
successfully sets the new value for flow span based on the flow pressure to
50.5 mmHg.
Send: set cal flow pres span 50.5
Receive: set cal flow pres span 50.5 ok
set cal flow span value
This command automatically sets the new value for flow span based on the
corrected flow entered. The following example sets the new value for flow
span based on the corrected flow to 17.2 LPM.