B-36 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
set baro mass coef value
This command sets the barometric mass coefficient to user-defined values
to value, where value is a floating-point representation of the coefficient.
The following example sets the barometric mass coefficient to 0.000630.
Send: set baro mass coef 0.000630
Receive: set baro mass coef 0.000630 ok
baro span
This command reports the current barometric pressure span recorded at the
time of calibration. The following example reports that the barometric span
is 1.0.
Send: baro span
Receive: baro span 1.0000
set baro span value
This command requires to manually specifying the span value for the
barometric pressure span, where value is a floating-point representation of
barometric pressure span. The following example successfully sets the
barometric span to 1.023.
Send: set baro span 1.023
Receive: set baro span 1.023 ok
beta th
This command reports the current reference beta threshold for the
detector. The following example reports that the beta threshold is 440.
Send: beta th
Receive: beta th 440
set beta th value
This command sets the reference beta threshold for the detector to value,
where value is a floating-point representation of the beta threshold. The
following example sets the beta threshold to 500.
Send: set beta th 500
Receive: set beta th 500 ok
bkg pm
This command reports the current PM backgrounds. The following
example reports that the PM background is 4 μg/m
3
.
Send: bkg pm
Receive: bkg pm 4.000E+00 μg/m3