Instruction manual
C-Link Protocol Commands
Calibration
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-35
set alpha th value
This command sets the alpha threshold for the detector to value, where
value is a floating-point number representing the alpha threshold for the
detector. The following example sets the alpha threshold to 1250.
Send: set alpha th 1250
Receive: set alpha th 1250 ok
amb rh offset
This command reports the current ambient RH offset, in percent. The
following example reports that the ambient RH offset is 0.6%.
Send: amb rh offset
Receive: amb rh offset 0.6 %
set amb rh offset value
This command sets the ambient RH offset to value, where value is a
floating-point number representing ambient RH offset in percent. The
following example sets the ambient RH offset to 7%.
Send: set amb rh offset 7
Receive: set amb rh offset 7 ok
amb temp offset
This command reports the current ambient temperature offset, in degrees
C. The following example reports that the ambient temperature offset is 5
°C.
Send: amb temp offset
Receive: amb temp offset 5.0 degC
set amb temp offset value
This command sets the ambient temperature offset to value, where value is
a floating-point number representing ambient temperature offset in degrees
C. The following example sets the ambient temperature offset to 3 °C.
Send: set amb temp offset 3
Receive: set amb temp offset 3 ok
baro mass coef
This command reports the current barometric mass coefficient. The
following example reports that the barometric mass coefficient is 0.000540.
Send: baro mass coef
Receive: baro mass coef 0.000540