Instruction manual
C-Link Protocol Commands
Calibration
B-34 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
stream time
This command reports if the streaming data string will have a time stamp
attached to it or not, according to Table B–4. The following example
reports that the streaming data shall not include a time stamp.
Send: stream time
Receive: stream time 0
set stream time value
This command enables value, where value is to attach or disable time stamp
to streaming data string, according to Table B–4. The following example
attaches a time stamp to streaming data.
Send: set stream time 1
Receive: set stream time 1 ok
Table B–4. Stream Time Values
Value Stream Time
0 Disables time stamp to streaming data string
1 Attaches time stamp to streaming data string
alpha eff
This command reports the current value of alpha efficiency for the
detector. The following example reports that the alpha efficiency is 0.12.
Send: alpha eff
Receive: alpha eff 0.120
set alpha eff value
This command sets the alpha efficiency for the detector to value, where
value is a floating-point number representing the alpha efficiency for the
detector. The following example sets the alpha efficiency to.0.15.
Send: set alpha eff 0.15
Receive: set alpha eff 0.15 ok
alpha th
This command reports the current value of alpha threshold for the
detector. The following example reports that the alpha threshold is 1220.
Send: alpha th
Receive: alpha th 1220
Calibration